## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K974025 · decision 1998-07-06 · decision SESE · status: Active

## Recalls

- Z-1940-2017 · Class II · terminated · initiated 2016-10-06: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1959-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1960-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1961-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1962-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1966-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1978-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1981-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1982-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1983-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1984-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-1989-2017 · Class II · terminated · initiated 2016-10-06 · matched through shared submission K974025, may concern other models: Supplied Directions for Use (DFU) did not contain a recommended service interval.
- Z-2077-2012 · Class II · terminated · initiated 2011-09-19 · matched through shared submission K974025, may concern other models: Anspach Effort, Inc. Palm Beach Gardens, FL is recalling their Single Use Sterile Bone Cutting Burrs used with XMax, microMax, microMax Plus, eMax, eMax2 and eMax2 Plus High Speed Drill Systems. It is possible for the cutter to come apart.
- Z-0829-2014 · Class II · terminated · initiated 2011-03-31 · matched through shared submission K974025, may concern other models: Cutter was longer than was stated on the label
- Z-0830-2014 · Class II · terminated · initiated 2011-03-31: Cutter was longer than was stated on the label
- Z-3202-2011 ·  · terminated · initiated 2011-03-17 · matched through shared submission K974025, may concern other models: The Anspach Effort, Palm Beach Gardens, FL, is recalling their Single Use, Sterile Bone Cutting Burrs, Catalog #, S-1504TD, Lot #s D363039140. D323038110, D293037349 and Catalog # TURQ-434-4, Lot #s D343038845, D323038131, due to the burr being 2.0 mm longer than stated on the label. As a result, a hole drilled using this burr would be 2 mm deeper than the screw.
- Z-3203-2011 ·  · terminated · initiated 2011-03-17 · matched through shared submission K974025, may concern other models: The Anspach Effort, Palm Beach Gardens, FL, is recalling their Single Use, Sterile Bone Cutting Burrs, Catalog #, S-1504TD, Lot #s D363039140. D323038110, D293037349 and Catalog # TURQ-434-4, Lot #s D343038845, D323038131, due to the burr being 2.0 mm longer than stated on the label. As a result, a hole drilled using this burr would be 2 mm deeper than the screw.

## Identifiers

- 510(k) Number: K974025 [FDA 510(k)]
- FDA product code: GFF [FDA 510(k)]

## Classification

- Device class: Class I [FDA 510(k)]
- Regulation: 21 CFR Part 878.4820 [FDA 510(k)]
- Advisory committee: General, Plastic Surgery [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Anspach Shielded Attachment, The Anspach Effort, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/anspach-shielded-attachment-gff (data as of 2026-11-28).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-28
- JSON: <https://devicerts.com/devices/anspach-shielded-attachment-gff.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/anspach-shielded-attachment-gff
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.