Antimicrobial Multi-Lumen Central Venous Catheter
510(k) K900263
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESK | NumberK900263 | Date24 Jul 1990 | StatusActive |
Recalls
-
- Ongoing Z-2376-2026
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
-
- Terminated Z-0053-2020
Lidstock contains a labeling error. The lidstock shows the catheter cross section of the gauge (GA) values in a reversed manner.
-
- Terminated Z-2886-2018
One lot of product is from product code AK-22502 so it contains a 2-lumen, 5 Fr x 5 cm catheter. However, this product is incorrectly labeled as AK-25502, so the lidstock says it contains a 2-lumen, 5 Fr x 8 cm catheter. No patient injuries have been reported related to this issue.
-
- Terminated Z-1984-2018
- Terminated Z-1985-2018
Product sterility may be compromised due to unsealed packaging.
-
- Terminated Z-1333-2018
Certain lots of cardiovascular catheters may not be completely sealed and so sterility of the product cannot be guaranteed.
-
- Terminated Z-1001-2018
Some of the products listed in the recall may be missing a sticker that contains the lot number, expiration date, and Unique Device Identification (UDI) information.
-
- Terminated Z-2337-2017
The Arrow CVC is indicated to provide short-term (<30 days) central venous access for treatment of diseases or conditions requiring central venous access including, but not limited to: -multiple infusions of fluids, medications, or chemotherapy -infusion of fluids that are hypertonic, hyperosmolar, or have divergent pH values -frequent blood sampling or blood/blood components infusions -infusion of incompatible medications -central venous pressure monitoring lack of usable peripheral IV sites -replacement of multiple peripheral sites for IV access
-
- Terminated Z-1830-2012
Lidstock of product does not contain chlorhexidine contraindication and contains wording "contains no medication" on the label. Product insert does reflect medicated status.
-
- Terminated Z-2496-2010
Incorrectly assembled 18 Ga x 2 1/2 Introducer Needle
-
- Terminated Z-1499-2009
Leakage: Material not lipid resistant, may crack and/or leak. Use of this device with lipid containing solutions or 70% isopropyl alcohol may cause leakage or air embolism.
-
- Terminated Z-0082-04
wrong product in box
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FOZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5200
- Advisory Committee
- General Hospital
- FDA Decision
- SESK Substantially Equivalent — Kit