Aortic Arch Cannulas
by Sorin Group Italia S.R.L. United States 510(k)
510(k) K943934 Product code DWF
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK943934 | Date2 Jun 1995 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2871-2017
Identification of excess plastic on the tip of the cannula.
-
- Terminated Z-2756-2016
Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.
-
- Terminated Z-2147-2016
Sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of approximately 45 degrees.
-
- Terminated Z-0813-2016
Sorin Group initiated field correction for the Dual Stage Venous Return Catheter because the incorrect Instructions For Use (IFU) was packaged in the case.
Specification
- 510(k) Number
- FDA Product Code
- DWF
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.4210
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent
Other devices from Sorin Group Italia S.R.L.
Properties & raw data
Show properties, registry sources and raw records
Sources (1)
- FDA 510(k) K943934