Skip to content

Apexpro Telemetry System

by General Electric Medical Systems Information Techn

510(k) K000779

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK000779 Date18 May 2000ClearedStatusActive

Recalls

2 recalls · 1 ongoing · FDA

    • Ongoing Z-0175-2020 FDA Class II initiated 30 Aug 2019

    May not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions.

    • Terminated Z-0725-2020 FDA Class II initiated 12 Nov 2019

    When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DSI
Device Class
Class II
Regulation
21 CFR Part 870.1025
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from General Electric Medical Systems Information Techn