Apollo Gamma Camera System

by Adac Laboratories

510(k) K062298

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK062298 Date23 Aug 2006ClearedStatusActive

Recalls

27 recalls · 2 ongoing · FDA

    • Ongoing Z-0929-2024 FDA Class I initiated 15 Dec 2023

    Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

    • Ongoing Z-0677-2024 FDA Class II initiated 30 Nov 2023

    While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.

  • 3 recalls, same reason

    • Terminated Z-0510-2018 FDA Class II initiated 15 Dec 2017
    • Terminated Z-0511-2018 FDA Class II initiated 15 Dec 2017
    • Terminated Z-0512-2018 FDA Class II initiated 15 Dec 2017

    During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.

  • 4 recalls, same reason

    • Terminated Z-1750-2017 FDA Class II initiated 24 Feb 2017
    • Terminated Z-1752-2017 FDA Class II initiated 24 Feb 2017
    • Terminated Z-1753-2017 FDA Class III initiated 24 Feb 2017
    • Terminated Z-1751-2017 FDA Class II initiated 24 Feb 2017

    The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.

  • 3 recalls, same reason

    • Terminated Z-1480-2017 FDA Class II initiated 31 Jan 2017
    • Terminated Z-1481-2017 FDA Class II initiated 31 Jan 2017
    • Terminated Z-1479-2017 FDA Class II initiated 31 Jan 2017

    Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan

    • Terminated Z-2735-2016 FDA Class II initiated 25 May 2016

    The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.

  • 2 recalls, same reason

    • Terminated Z-1021-2016 FDA Class II initiated 17 Dec 2015
    • Terminated Z-1020-2016 FDA Class II initiated 17 Dec 2015

    The firm discovered four issues with the BrightView Family systems. Issue 1: During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with the bottom edge of the patient pallet. If this occurs, collision sensors would pause the scan. Issue 2: During a scan, the gantry motion may halt prematurely. This cause

  • 2 recalls, same reason

    • Terminated Z-1791-2015 FDA Class II initiated 6 Mar 2015
    • Terminated Z-1790-2015 FDA Class II initiated 6 Mar 2015

    The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.

    • Terminated Z-0450-2015 FDA Class II initiated 4 Nov 2014

    Software issues

    • Terminated Z-1481-2013 FDA Class II initiated 25 Apr 2013

    Philips has received one report from the field that there was an unexpected motion of the detector due to a failure of the mechanical assembly.

    • Terminated Z-2947-2011 FDA Unclassified (pre-2012) initiated 15 Jun 2011

    Non-conformances with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly.

    • Terminated Z-2867-2011 FDA Unclassified (pre-2012) initiated 3 Jun 2011 on shared 510(k) K062298

    The gap between the two parts of the table may permit pinches to occur on either patients skin or other body parts; it has the potential for serious injury if a patients finger is caught between the pallet and patient table resulting in broken finger bone.

  • 5 recalls, same reason

    • Terminated Z-2870-2011 FDA Unclassified (pre-2012) initiated 3 Jun 2011
    • Terminated Z-2868-2011 FDA Unclassified (pre-2012) initiated 3 Jun 2011
    • Terminated Z-2872-2011 FDA Unclassified (pre-2012) initiated 3 Jun 2011
    • Terminated Z-2869-2011 FDA Unclassified (pre-2012) initiated 3 Jun 2011
    • Terminated Z-2871-2011 FDA Unclassified (pre-2012) initiated 3 Jun 2011

    The gap between the two parts of the table may permit pinches to occur on either patient's skin or other body parts; it has the potential for serious injury if a patient's finger is caught between the pallet and patient table resulting in broken finger bone.

    • Terminated Z-1892-2009 FDA Unclassified (pre-2012) initiated 2 Jul 2009

    Unexpected movement-- if the drive assembly is severely compromised, the detector may slide to its hardware limit.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
KPS
Device Class
Class II
Regulation
21 CFR Part 892.1200
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent