Apollo Gamma Camera System
510(k) K062298
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK062298 | Date23 Aug 2006 | StatusActive |
Recalls
-
- Ongoing Z-0929-2024
Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
-
- Ongoing Z-0677-2024
While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.
-
- Terminated Z-0510-2018
- Terminated Z-0511-2018
- Terminated Z-0512-2018
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.
-
- Terminated Z-1750-2017
- Terminated Z-1752-2017
- Terminated Z-1753-2017
- Terminated Z-1751-2017
The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
-
- Terminated Z-1480-2017
- Terminated Z-1481-2017
- Terminated Z-1479-2017
Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan
-
- Terminated Z-2735-2016
The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
-
- Terminated Z-1021-2016
- Terminated Z-1020-2016
The firm discovered four issues with the BrightView Family systems. Issue 1: During a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with the bottom edge of the patient pallet. If this occurs, collision sensors would pause the scan. Issue 2: During a scan, the gantry motion may halt prematurely. This cause
-
- Terminated Z-1791-2015
- Terminated Z-1790-2015
The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
-
- Terminated Z-0450-2015
Software issues
-
- Terminated Z-1481-2013
Philips has received one report from the field that there was an unexpected motion of the detector due to a failure of the mechanical assembly.
-
- Terminated Z-2947-2011
Non-conformances with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly.
-
- Terminated Z-2867-2011
The gap between the two parts of the table may permit pinches to occur on either patients skin or other body parts; it has the potential for serious injury if a patients finger is caught between the pallet and patient table resulting in broken finger bone.
-
- Terminated Z-2870-2011
- Terminated Z-2868-2011
- Terminated Z-2872-2011
- Terminated Z-2869-2011
- Terminated Z-2871-2011
The gap between the two parts of the table may permit pinches to occur on either patient's skin or other body parts; it has the potential for serious injury if a patient's finger is caught between the pallet and patient table resulting in broken finger bone.
-
- Terminated Z-1892-2009
Unexpected movement-- if the drive assembly is severely compromised, the detector may slide to its hardware limit.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KPS
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent