Apollo Revision/Constrained Knee System
510(k) K972501
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK972501 | Date1 Oct 1997 | StatusActive |
Recalls
-
- Terminated Z-1261-2014
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JWH
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3560
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent