Aq Hydrophilic Stent
by Cook Urological, Inc. 510(k)
Active
510(k) K961446
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK961446 | Date28 May 1996 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0977-2020
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
-
- Terminated Z-0852-2011
The firm received a customer complaint that a box labeled as Universa Firm Stents but the box actually contained Universa Soft Stents. The inner product label is correct but the outer box is wrong.
-
- Terminated Z-0980-2009
"Soft" stents were placed in "firm" stent outer boxes. The soft stent is only to be left in the patient for 6 months, but the firm stent can be left in the patient for 12 months, so this error could result in the stent being left in the patient for longer than indicated for the device.
-
- Terminated Z-0839-2009
- Terminated Z-0841-2009
- Terminated Z-0850-2009
- Terminated Z-0842-2009
- Terminated Z-0844-2009
- Terminated Z-0843-2009
- Terminated Z-0845-2009
- Terminated Z-0849-2009
- Terminated Z-0846-2009
- Terminated Z-0853-2009
- Terminated Z-0851-2009
- Terminated Z-0848-2009
- Terminated Z-0847-2009
- Terminated Z-0852-2009
- Terminated Z-0840-2009
The bond may break where the two materials meet, resulting in separation of the stent.
Specification
- 510(k) Number
- FDA Product Code
- FAD
- Device Class
- Class II
- Regulation
- 21 CFR Part 876.4620
- Advisory Committee
- Gastroenterology, Urology
- FDA Decision
- SESE Substantially Equivalent