  "format": "devicerts.device.v1",
  "cite_as": "ARIES System, Luminex Corporation. deviCERTS medical device catalogue. https://devicerts.com/devices/aries-system-ooi (data as of 2026-11-20).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/aries-system-ooi",
    "json_url": "https://devicerts.com/devices/aries-system-ooi.json",
    "markdown_url": "https://devicerts.com/devices/aries-system-ooi.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-20",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/aries-system-ooi",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "ARIES System",
    "registry_name": "ARIES System",
    "url": "https://devicerts.com/devices/aries-system-ooi",
    "manufacturer": {
      "name": "Luminex Corporation",
      "url": "https://devicerts.com/manufacturers/luminex-corporation",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K160517",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "OOI"
    ],
    "emdn": [],
    "gmdn": [],
    "description": {
      "text": "The Aries System is a Real-Time Nucleic Acid Amplification System, designed for molecular diagnostics in clinical chemistry. It automates nucleic acid extraction, real-time PCR amplification, and signal detection for multiplex assays, enabling simultaneous analysis of multiple targets from various sample types.\n\nThis system is intended for use in clinical laboratory settings and is supplied as a reusable device. It operates under sterile conditions for nucleic acid processing and detection, with no specific MRI safety restrictions noted. The system supports multiple assay configurations, as indicated by its regulatory registration numbers, and holds FDA 510(k) clearance under Special Review (K160517). Sterility and maintenance protocols are manufacturer-defined, with no explicit single-use designation.",
      "ai_generated": true,
      "generated_on": "2026-09-18",
      "note": "Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data is the record."
    },
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K160517",
      "number_label": null,
      "device_udi_di": null,
      "date": "2016-04-12",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2288-2019",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2019-02-20",
      "reason": "Through internal investigation, it was discovered that  a portion  of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.",
      "applies": "this device",
      "via": "K160517"
    },
    {
      "number": "Z-2292-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-06-29",
      "reason": "An incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient order.",
      "applies": "this device",
      "via": "K160517"
    },
    {
      "number": "Z-2094-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-04-19",
      "reason": "The hand-held barcode scanner that is used as an accessory to enter ARIES cassette and specimen identifications on the ARIES Systems populates patient sample IDs with a previously scanned sample ID when entering orders into an ARIES instrument.",
      "applies": "this device",
      "via": "K160517"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K160517",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "OOI",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 862.2570",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Clinical Chemistry",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}