Arietta 750
by Hitachi Healthcare Americas Corporation
510(k) K191233
Active
510(k)
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK191233 | Date9 Aug 2019 | StatusActive |
Recalls
-
- Ongoing Z-0189-2022
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IYN
- FDA applicant
- Hitachi Healthcare Americas
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1550
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Other devices from Hitachi Healthcare Americas Corporation
See all devices from Hitachi Healthcare Americas Corporation →