Arrow IPN000067
Arrow IPN000067 (AI-07121) is a temporary cardiac pacing balloon catheter (FDA Class II; FDA product code DYG, Catheter, Flow Directed) made by Teleflex Incorporated, cleared under FDA 510(k) K892530 on 26 Sep 1989 and listed in FDA GUDID since 16 Sep 2016 (UDI-DI 00801902002822), active as of 22 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00801902002822 | Date16 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK892530 | Date26 Sep 1989 | StatusActive |
Official records
- AccessGUDID record 00801902002822
- FDA 510(k) database K892530
From the FDA record
- Device description
- 4 Fr. 60 cm Balloon Wedge Pressure Catheter
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- Labelled as containing natural rubber latex
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K892530 decided 26 Sep 1989 Active
- FDA GUDID UDI-DI 00801902002822 listed 16 Sep 2016 Listed
Recalls
-
- Terminated Z-0715-2020
- Terminated Z-0716-2020
Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death
-
- Terminated Z-1057-2013
The product specifications sheet, part of the Instructions for Use (IFU), for the specified product numbers, reference Renografin-76 contrast medium (viscosity of 8.4 centipoise), which is no longer available. Firm is in the process of updating the labeling for these products. In the meantime, when choosing an alternative contrast medium, health care providers are instructed to select a contrast
Timeline
- Cleared FDA 510(k) K892530
- Recall initiated FDA recalls Z-1057-2013
- Listed FDA GUDID 00801902002822
- Recall initiated FDA recalls Z-0715-2020
- Recall initiated FDA recalls Z-0716-2020
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Arrow
- Generic name
- Temporary cardiac pacing balloon catheter
- Model / Ref / Catalog
- IPN000067, AI-07121
- Primary DI / UDI-DI
- 00801902002822
- 510(k) Number
- FDA product code
- DYG
- GMDN term
- Temporary cardiac pacing balloon catheter
- Device class
- Class II
- Regulation number
- 870.1240
- GUDID distribution status
- In Commercial Distribution
- Size
- Length — 60 CentimeterCatheter Gauge — 4 French
Similar devices
Other devices filed under FDA product code DYG (CATHETER, FLOW DIRECTED).
- Swan-Ganz Pacing Probe and Catheters; Swan-Ganz Bipolar Pacing Catheter (D97120F5); Swan-Ganz Bipolar Pacing Catheter (D97130F5); Swan-Ganz Pacing Thermodilution TD Catheter (D200F7); Chandler Transluminal V-Pacing Probe (D98100) by Edwards Lifesciences LLC
- VIGILANCE CONTINUOUS CARDIAC OUTPUT/OXIMETRY (CCO/SVO2) MONITOR by Baxter Healthcare Corp
- SWAN GANZ by Clincal Instruments Corp.
- Nova Balloon Thermodilution Catheter by Nova Medical Specialties
- HEPARIN COATED THERMODILUTION CATHETER by Kma, Inc.
- Medtronic Model 6550 Adjustable Hemostasis Valve by MEDTRONIC VASCULAR, INC.
- Silicone Flow-Directed Thermodilution Catheter by Cathlab Corp.
- Swan-Ganz Flow-Directed Thermodil Cath by American Edwards Laboratories
- Swan-Ganz Catheters by Edwards Lifesciences LLC
- Torque Control Balloon Catheter by Cordis US Corp.