Arrow IPN915248
Arrow IPN915248 (AI-07125) is a temporary cardiac pacing balloon catheter (FDA Class II; FDA product code DYG, Catheter, Flow Directed) made by Teleflex Incorporated, cleared under FDA 510(k) K892530 on 26 Sep 1989 and listed in FDA GUDID since 11 Dec 2019 (UDI-DI 10801902160543), active as of 23 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10801902160543 | Date11 Dec 2019 | StatusListed |
| United States | FDA 510(k)SESE | NumberK892530 | Date26 Sep 1989 | StatusActive |
Official records
- AccessGUDID record 10801902160543
- FDA 510(k) database K892530
From the FDA record
- Device description
- 6 Fr. 60 cm Balloon Wedge Pressure Catheter
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- Labelled as containing natural rubber latex
- Sterility
- Supplied sterile; sterilise before use (Ethylene Oxide)
Across registries
- FDA 510(k) K892530 decided 26 Sep 1989 Active
- FDA GUDID UDI-DI 10801902160543 listed 11 Dec 2019 Listed
Recalls
-
- Terminated Z-0715-2020
- Terminated Z-0716-2020
Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death
-
- Terminated Z-1057-2013
The product specifications sheet, part of the Instructions for Use (IFU), for the specified product numbers, reference Renografin-76 contrast medium (viscosity of 8.4 centipoise), which is no longer available. Firm is in the process of updating the labeling for these products. In the meantime, when choosing an alternative contrast medium, health care providers are instructed to select a contrast
Timeline
- Cleared FDA 510(k) K892530
- Recall initiated FDA recalls Z-1057-2013
- Recall initiated FDA recalls Z-0715-2020
- Recall initiated FDA recalls Z-0716-2020
- Listed FDA GUDID 10801902160543
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Arrow
- Generic name
- Temporary cardiac pacing balloon catheter
- Model / Ref / Catalog
- IPN915248, AI-07125
- Primary DI / UDI-DI
- 10801902160543
- 510(k) Number
- FDA product code
- DYG
- GMDN term
- Temporary cardiac pacing balloon catheter
- Device class
- Class II
- Regulation number
- 870.1240
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Ethylene Oxide
- Size
- Catheter Gauge — 6 FrenchLength — 60 Centimeter
Similar devices
Other devices filed under FDA product code DYG (CATHETER, FLOW DIRECTED).
- LEONARD PERIPHERAL VENOUS PRESSURE CATH by Norton Performance Plastics Corp.
- THERMAL DILUTION CATHETER by Catheter Corp. of America
- FLOW DIRECT THERMODILUT BALLOON CATH HEPARIN COATED by J-Lloyd Medical, Inc.
- HemoSphere Alta Advanced Monitoring Platform (ALTAALL1/ALTCR1/ALTASR1); HemoSphere Alta- monitor pressure cable (HEMAPSC200);Acumen AFM cable- HemoSphere Alta monitor (HEMAFM100) by Edwards Lifesciences LLC
- INTERFLO MEDICAL MODEL PA3-H CCO/PAC CATHETER by Intec Medical, Inc.
- ELECATH BALTHERM TLC THERMAL DILUTION by Electro-Catheter Corp.
- OCCLUSION CATHETER by Bionomy, Inc.
- DUAL-THERMISTOR THERMODILUTION CATH. & INFU. CATH. by Nova Medical Specialties
- CONTINOUS CARDIAC OUTPUT PULMONARY ARTERY CATHETER, MODEL TRUCATH, CONTINOUS CARDIAC OUTPUT MONITOR, MODEL TRUCCOM by Alba Critical Care, Ltd.