# Arrow IPN037745

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/arrow-ask-05001-slr1>
- Manufacturer: [Teleflex Incorporated](https://devicerts.com/manufacturers/teleflex-incorporated) (FDA registration: United States)
- Name in the registry: ARROW IPN037745
- Device class: Class II (US FDA)
- FDA product code: CAZ
- GMDN: Epidural anaesthesia set, medicated (61911)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 10801902130966 · listed 2017-01-13 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K801912 · decision 1980-09-16 · decision SESE · status: Active · from GUDID

## Recalls

- Z-2364-2026 · Class I · ongoing · initiated 2026-04-21: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
- Z-2803-2020 · Class II · terminated · initiated 2020-07-06: The lidstock states the incorrect expiration date for the product. This issue could result in use of a device that is expired which could potentially lead to an increased risk of infection or other complications. Sterility, biocompatibility, safety, or efficacy of the kits and their components are not assured beyond the correct expiration date.
- Z-1175-2020 · Class II · terminated · initiated 2020-01-15: There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1249-2020 · Class II · terminated · initiated 2019-12-20: Product lidstock contains the incorrect expiration date for the product
- Z-1251-2020 · Class II · terminated · initiated 2019-12-20: Product lidstock contains the incorrect expiration date for the product
- Z-1246-2020 · Class II · terminated · initiated 2019-12-20: Product lidstock contains the incorrect expiration date for the product
- Z-1248-2020 · Class II · terminated · initiated 2019-12-20: Product lidstock contains the incorrect expiration date for the product
- Z-1247-2020 · Class II · terminated · initiated 2019-12-20: Product lidstock contains the incorrect expiration date for the product
- Z-1245-2020 · Class II · terminated · initiated 2019-12-20: Product lidstock contains the incorrect expiration date for the product
- Z-2012-2018 · Class II · terminated · initiated 2018-04-11: Product sterility may be compromised due to unsealed packaging.

## Identity

- Trade name: ARROW [FDA GUDID (AccessGUDID)]
- Generic name: Epidural anaesthesia set, medicated [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: IPN037745 [FDA GUDID (AccessGUDID)]; ASK-05001-SLR1 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 10801902130966 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K801912 [FDA GUDID (AccessGUDID)]
- FDA product code: CAZ [FDA GUDID (AccessGUDID)]
- GMDN term: Epidural anaesthesia set, medicated (61911) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 868.5140 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Clinical & Handling

- MRI safety: MR Unsafe [FDA GUDID (AccessGUDID)]

## Properties

- Supplied sterile: Yes [FDA GUDID (AccessGUDID)]
- Single use: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Convenience kit: Yes [FDA GUDID (AccessGUDID)]
- Combination product: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Expiration date: Yes [FDA GUDID (AccessGUDID)]
- Manufacturing date: Yes [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Arrow IPN037745, Teleflex Incorporated. deviCERTS medical device catalogue. https://devicerts.com/devices/arrow-ask-05001-slr1 (data as of 2026-11-23).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/arrow-ask-05001-slr1.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/arrow-ask-05001-slr1
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.