# Arrow IPN000206

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/arrow-iaa-00401>
- Manufacturer: [Teleflex Incorporated](https://devicerts.com/manufacturers/teleflex-incorporated) (FDA registration: United States)
- Name in the registry: ARROW IPN000206
- Device class: Class II (US FDA)
- FDA product code: DSP
- GMDN: Intra-aortic balloon pump (36340)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 00801902084552 · listed 2015-10-16 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K060309 · decision 2006-04-06 · decision SESE · status: Active · from GUDID

## Recalls

- Z-0426-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0427-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0428-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0429-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0430-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0431-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0432-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0444-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0445-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0446-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0447-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0448-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0449-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0450-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0451-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0452-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0453-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0454-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0455-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-0456-2023 · Class I · ongoing · initiated 2022-10-17: There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
- Z-2408-2020 · Class I · terminated · initiated 2020-05-20: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- Z-2409-2020 · Class I · terminated · initiated 2020-05-20: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- Z-2410-2020 · Class I · terminated · initiated 2020-05-20: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- Z-2411-2020 · Class I · terminated · initiated 2020-05-20: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- Z-2412-2020 · Class I · terminated · initiated 2020-05-20: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- Z-2413-2020 · Class I · terminated · initiated 2020-05-20: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- Z-2414-2020 · Class I · terminated · initiated 2020-05-20: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- Z-2415-2020 · Class I · terminated · initiated 2020-05-20: A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
- Z-0550-2008 ·  · terminated · initiated 2007-11-12: Leak may occur in the helium drive system for the IABP

## Identity

- Trade name: ARROW [FDA GUDID (AccessGUDID)]
- Generic name: Intra-aortic balloon pump [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: IPN000206 [FDA GUDID (AccessGUDID)]; IAA-00401 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 00801902084552 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K060309 [FDA GUDID (AccessGUDID)]
- FDA product code: DSP [FDA GUDID (AccessGUDID)]
- GMDN term: Intra-aortic balloon pump (36340) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 870.3535 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Clinical & Handling

- MRI safety: MR Unsafe [FDA GUDID (AccessGUDID)]

## Properties

- Supplied sterile: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Arrow IPN000206, Teleflex Incorporated. deviCERTS medical device catalogue. https://devicerts.com/devices/arrow-iaa-00401 (data as of 2026-11-29).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-29
- JSON: <https://devicerts.com/devices/arrow-iaa-00401.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/arrow-iaa-00401
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.