Skip to content

Arrow IPN915854

by Teleflex Incorporated

Ref IAA-00437Model IPN915854

Arrow IPN915854 (IAA-00437) is an intra-aortic balloon pump (EU MDD Class I; FDA Class II; FDA product code DSP, System, balloon, intra-aortic and control) made by Teleflex Incorporated, registered in EUDAMED on 4 Aug 2026 (Basic UDI-DI B-00801902172396), cleared under FDA 510(k) K060309 on 6 Apr 2006 and listed in FDA GUDID since 16 Sep 2021 (UDI-DI 00801902172396), active as of 22 Oct 2026. 29 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE MDD 510(k)

Valid until

No date given

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

20 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

510(k) K060309

FDA submission, from GUDID

Trust

Unverified

Source EUDAMED, GUDID, synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDD EUDAMED NumberB-00801902172396 Basic UDI-DI 00801902172396UDI-DIDate4 Aug 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number00801902172396 UDI-DI Date16 Sep 2021ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE GUDID from GUDID Substantially Equivalent · Special NumberK060309 Date6 Apr 2006DecidedStatusActive

Official records

From the FDA record

MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
Labelled as not made with natural rubber latex
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 22 Oct 2026, as the labeller entered it.

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI B-00801902172396 registered 4 Aug 2026 Active
  • FDA 510(k) K060309 decided 6 Apr 2006 Active
  • FDA GUDID UDI-DI 00801902172396 listed 16 Sep 2021 Listed

Recalls

29 recalls · 20 ongoing · FDA

  • 20 recalls, same reason

    • Ongoing Z-0426-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0427-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0428-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0429-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0430-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0431-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0432-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0444-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0445-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0446-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0447-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0448-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0449-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0450-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0451-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0452-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0453-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0454-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0455-2023 FDA Class I initiated 17 Oct 2022
    • Ongoing Z-0456-2023 FDA Class I initiated 17 Oct 2022

    There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.

  • 8 recalls, same reason

    • Terminated Z-2408-2020 FDA Class I initiated 20 May 2020
    • Terminated Z-2409-2020 FDA Class I initiated 20 May 2020
    • Terminated Z-2410-2020 FDA Class I initiated 20 May 2020
    • Terminated Z-2411-2020 FDA Class I initiated 20 May 2020
    • Terminated Z-2412-2020 FDA Class I initiated 20 May 2020
    • Terminated Z-2413-2020 FDA Class I initiated 20 May 2020
    • Terminated Z-2414-2020 FDA Class I initiated 20 May 2020
    • Terminated Z-2415-2020 FDA Class I initiated 20 May 2020

    A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

    • Terminated Z-0550-2008 FDA Unclassified (pre-2012) initiated 12 Nov 2007

    Leak may occur in the helium drive system for the IABP

Timeline

32 dated events

  1. Recall initiated FDA recalls Z-0448-2023
  2. Recall initiated FDA recalls Z-0449-2023
  3. Recall initiated FDA recalls Z-0450-2023
  4. Recall initiated FDA recalls Z-0451-2023
  5. Recall initiated FDA recalls Z-0452-2023
  6. Recall initiated FDA recalls Z-0453-2023
  7. Recall initiated FDA recalls Z-0454-2023
  8. Recall initiated FDA recalls Z-0455-2023
  9. Recall initiated FDA recalls Z-0456-2023
  10. Registered EUDAMED B-00801902172396

22 earlier events are in the sections above.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDD Active
  • United States FDA 510(k) Active

Specification

EUDAMED · GUDID

Trade name
5 Lead ECG Cable Assembly AHA For use with IABP EUDAMED
Arrow GUDID
Generic name
Intra-aortic balloon pump
Device name
5 Lead ECG Cable Assembly Aha
Model / Ref / Catalog
IPN915854, IAA-00437
Primary DI / UDI-DI
00801902172396
Basic UDI-DI
B-00801902172396
510(k) Number
FDA product code
DSP
EMDN code
Z12059099 Various instruments for cardiology and cardiac surgery -other
GMDN term
Intra-aortic balloon pump
Device class
Class I EUDAMED
Class II GUDID
Regulation number
870.3535
EU registration type
Legacy device (under a directive)
EU market of registration
Ireland
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device classEUDAMED
Class I
GUDID
Class II

Device class is stated on each registry's own scale: Class I in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Similar devices in this category

Other devices filed under VARIOUS INSTRUMENTS FOR CARDIOLOGY AND CARDIAC SURGERY -OTHER.

Other devices from Teleflex Incorporated

  • Arrow IPNB00820 Open recall

    Class II
  • Arrow IPNB00826 Open recall

    Class II
  • Arrow IPNB00822 Open recall

    Class II
  • Arrow IPNB00827 Open recall

    Class II

See all devices from Teleflex Incorporated →

Properties

  • Prescription use
Other properties (24, all answered No)