Arrow IPN000255
Arrow IPN000255 (IAB-05840-U) is an intra-aortic balloon catheter (FDA Class II; FDA product code DSP, System, balloon, intra-aortic and control) made by Teleflex Incorporated, cleared under FDA 510(k) K000729 on 19 May 2000 and listed in FDA GUDID since 16 Oct 2015 (UDI-DI 00801902010759), active as of 22 Oct 2026. 27 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00801902010759 | Date16 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK000729 | Date19 May 2000 | StatusActive |
Official records
- AccessGUDID record 00801902010759
- FDA 510(k) database K000729
Across registries
- FDA 510(k) K000729 decided 19 May 2000 Active
- FDA GUDID UDI-DI 00801902010759 listed 16 Oct 2015 Listed
Recalls
-
- Ongoing Z-1879-2024
- Ongoing Z-1880-2024
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.
-
- Terminated Z-1059-2016
- Terminated Z-1062-2016
- Terminated Z-1060-2016
- Terminated Z-1058-2016
- Terminated Z-1056-2016
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.
-
- Terminated Z-1064-2013
- Terminated Z-1065-2013
Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
-
- Terminated Z-0792-2011
- Terminated Z-0969-2011
- Terminated Z-0790-2011
- Terminated Z-0793-2011
- Terminated Z-0791-2011
SuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.
-
- Terminated Z-1215-2009
- Terminated Z-1222-2009
- Terminated Z-1221-2009
- Terminated Z-1220-2009
- Terminated Z-1214-2009
- Terminated Z-1213-2009
- Terminated Z-1211-2009
- Terminated Z-1217-2009
- Terminated Z-1219-2009
- Terminated Z-1218-2009
- Terminated Z-1216-2009
- Terminated Z-1210-2009
- Terminated Z-1212-2009
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Timeline
- Recall initiated FDA recalls Z-1064-2013
- Recall initiated FDA recalls Z-1065-2013
- Listed FDA GUDID 00801902010759
- Recall initiated FDA recalls Z-1059-2016
- Recall initiated FDA recalls Z-1062-2016
- Recall initiated FDA recalls Z-1060-2016
- Recall initiated FDA recalls Z-1058-2016
- Recall initiated FDA recalls Z-1056-2016
- Recall initiated FDA recalls Z-1879-2024
- Recall initiated FDA recalls Z-1880-2024
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Arrow
- Generic name
- Intra-aortic balloon catheter
- Model / Ref / Catalog
- IPN000255, IAB-05840-U
- Primary DI / UDI-DI
- 00801902010759
- 510(k) Number
- FDA product code
- DSP
- GMDN term
- Intra-aortic balloon catheter
- Device class
- Class II
- Regulation number
- 870.3535
- GUDID distribution status
- In Commercial Distribution
- Size
- Total Volume — 40 Milliliter
Similar devices
Other devices filed under FDA product code DSP (System, balloon, intra-aortic and control).
- ARROW KINK RESISTANT FLEXI-CATH INTRA-AORTIC BALLOON by Arrow Intl., Inc.
- SUPRACOR INTR-AORTIC BALLOON PUMP CONSOLE by Abiomed, Inc.
- SUPRACOR BALLOON CATHETER by Abiomed, Inc.
- 8 FR PERCUTAN DOUBLE LUMEN INTRA-AORTIC BALLOON by Kontron Instruments, Inc.
- 40CC PERCUTANEOUS PRE-FURL (D/L) BALLOON CATHETER by Aries Medical, Inc.
- ACAT 1 (ARROW CARDIAC ASSIST TECHNOLOGY 1) by Arrow Intl., Inc.
- CARDIOSAVE HYBRID TYPE D PLUG 0998-00-0800-33 by Datascope Corp.
- PRESSURE TRIGGER MOD OF SMEC INTRA-AORT by Smec, Inc.
- SIDEWINDER 40CC DL INTRA-AORTIC BALLOON CATHETER by Boston Scientific
- PECOR, PERCOR-DL, 10.5 FRENCH W/SMEC by Datascope Corp.