Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement System, Models Iab-05830-F, IAB5840
510(k) K021462
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK021462 | Date6 Jun 2002 | StatusActive |
Recalls
-
- Terminated Z-1055-2016
- Terminated Z-1057-2016
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.
-
- Terminated Z-1233-2015
Outer package product label incorrectly identifies the product as FiberOptix Flex IAB 7.5Fr rather than the correct FiberOptix IAB 8Fr
-
- Terminated Z-1259-2015
Potential for a leak between the catheter and the bifurcate which would allow helium to escape to the atmosphere.
-
- Terminated Z-1065-2013
- Terminated Z-1064-2013
Potential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
-
- Terminated Z-0790-2011
- Terminated Z-0792-2011
- Terminated Z-0969-2011
- Terminated Z-0793-2011
- Terminated Z-0791-2011
SuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.
-
- Terminated Z-1222-2009
- Terminated Z-1212-2009
- Terminated Z-1221-2009
- Terminated Z-1220-2009
- Terminated Z-1214-2009
- Terminated Z-1213-2009
- Terminated Z-1211-2009
- Terminated Z-1217-2009
- Terminated Z-1219-2009
- Terminated Z-1218-2009
- Terminated Z-1216-2009
- Terminated Z-1210-2009
- Terminated Z-1215-2009
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DSP
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.3535
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent