Arrow IPN048022
by Teleflex Incorporated 2 variants
Arrow IPN048022 (OT19TCGLK) is an anaesthesia conduction catheter (FDA Class II; FDA product code BSO, Catheter, Conduction, Anesthetic) made by Teleflex Incorporated, cleared under FDA 510(k) K140110 on 27 Mar 2015 and listed in FDA GUDID since 8 Jul 2017 (UDI-DI 10801902130645), active as of 23 Oct 2026. 9 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10801902130645 | Date8 Jul 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK140110 | Date27 Mar 2015 | StatusActive |
Official records
- AccessGUDID record 10801902130645
- FDA 510(k) database K140110
From the FDA record
- Device description
- Arrow Flex Tip Plus Epidural Catherization Kit with NRFit(R) connector
- MRI safety
- MR Conditional
- Natural rubber latex
- Labelled as not made with natural rubber latex
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K140110 decided 27 Mar 2015 Active
- FDA GUDID UDI-DI 10801902130645 listed 8 Jul 2017 Listed
Recalls
-
- Ongoing Z-2371-2026
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
-
- Ongoing Z-1913-2026
Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured
-
- Ongoing Z-2288-2020
Nine complaints reported that the rotatable collar on the filter had detached, causing a leak.
-
- Terminated Z-1174-2020
- Terminated Z-1177-2020
- Terminated Z-1172-2020
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
-
- Terminated Z-1238-2020
- Terminated Z-1236-2020
- Terminated Z-1237-2020
Product lidstock contains the incorrect expiration date for the product
Timeline
- Listed FDA GUDID 10801902130645
- Recall initiated FDA recalls Z-1238-2020
- Recall initiated FDA recalls Z-1236-2020
- Recall initiated FDA recalls Z-1237-2020
- Recall initiated FDA recalls Z-1174-2020
- Recall initiated FDA recalls Z-1177-2020
- Recall initiated FDA recalls Z-1172-2020
- Recall initiated FDA recalls Z-2288-2020
- Recall initiated FDA recalls Z-1913-2026
- Recall initiated FDA recalls Z-2371-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Arrow
- Generic name
- Anaesthesia conduction catheter
- Model / Ref / Catalog
- IPN048022, OT19TCGLK
- Primary DI / UDI-DI
- 10801902130645
- 510(k) Number
- FDA product code
- BSO
- GMDN term
- Anaesthesia conduction catheter
- Device class
- Class II
- Regulation number
- 868.5120
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Conditional
Similar devices
Other devices filed under FDA product code BSO (CATHETER, CONDUCTION, ANESTHETIC).
- 4 SURE TM ANESTHESIA EXTENSION SET by Boston Pacific Medical, Inc.
- C-CAT Anaesthesia Catheter kit by Cimpax Aps
- BUSSE EPIDURAL CATHETER KIT by BUSSE HOSPITAL DISPOSABLES
- Medline Reinforced Epidural Catheter by MEDLINE INDUSTRIES, INC.
- Perifix FX Catheter; Contiplex FX Catheter by B. Braun Medical, Inc.
- ON-Q Pain Relief System, QuikBloc Over-the-Needle Catheter Set by KIMBERLY-CLARK CORPORATION
- Medline Epidural Catheter by MEDLINE INDUSTRIES, INC.
- EPINAUT by IMEDICOM CO., LTD.
- EVANS, VAS-CATH EPIDURAL NEEDLE & CATHETER by Vas-Cath of Canada, Ltd.
- Certa Catheter by Ferrosan Medical Devices Sp. Z.O.O.