Skip to content

Arrow-Trerotola Percutaneous Thrombolytic Device or Ptd

by Arrow Intl., Inc.

510(k) K011056

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

13 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK011056 Date24 Oct 2001ClearedStatusActive

Recalls

13 recalls · FDA

  • 7 recalls, same reason

    • Terminated Z-0624-2022 FDA Class I initiated 11 Jan 2022
    • Terminated Z-0629-2022 FDA Class I initiated 11 Jan 2022
    • Terminated Z-0625-2022 FDA Class I initiated 11 Jan 2022
    • Terminated Z-0626-2022 FDA Class I initiated 11 Jan 2022
    • Terminated Z-0628-2022 FDA Class I initiated 11 Jan 2022
    • Terminated Z-0623-2022 FDA Class I initiated 11 Jan 2022
    • Terminated Z-0627-2022 FDA Class I initiated 11 Jan 2022

    The firm received reports indicating PTD tip separation during use.

  • 4 recalls, same reason

    • Terminated Z-0231-2022 FDA Class I initiated 20 Sep 2021 on shared 510(k) K011056
    • Terminated Z-0233-2022 FDA Class I initiated 20 Sep 2021 on shared 510(k) K011056
    • Terminated Z-0234-2022 FDA Class I initiated 20 Sep 2021 on shared 510(k) K011056
    • Terminated Z-0232-2022 FDA Class I initiated 20 Sep 2021 on shared 510(k) K011056

    7FR Percutaneous Thrombolytic Device (PTD) Inner lumen (orange) detached from the device basket result in embolization and retrieval with an additional intervention and consequent delay of therapy.

    • Terminated Z-1502-2009 FDA Unclassified (pre-2012) initiated 11 Feb 2009 on shared 510(k) K011056

    Leakage: Material not lipid resistant, may crack and/or leak. Use of this device with lipid containing solutions or 70% isopropyl alcohol may cause leakage or air embolism.

    • Terminated Z-0842-05 FDA Unclassified (pre-2012) initiated 18 Mar 2005 on shared 510(k) K011056

    The orange Pebax lumen located within the catheter basket may separate from the catheter basket sleeve permitting the lumen to become dislodged during use.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
QEW
Device Class
Class II
Regulation
21 CFR Part 870.5150
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Arrow Intl., Inc.

See all devices from Arrow Intl., Inc. →