Arrow-Trerotola Percutaneous Thrombolytic Device or Ptd
510(k) K011056
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK011056 | Date24 Oct 2001 | StatusActive |
Recalls
-
- Terminated Z-0624-2022
- Terminated Z-0629-2022
- Terminated Z-0625-2022
- Terminated Z-0626-2022
- Terminated Z-0628-2022
- Terminated Z-0623-2022
- Terminated Z-0627-2022
The firm received reports indicating PTD tip separation during use.
-
- Terminated Z-0231-2022
- Terminated Z-0233-2022
- Terminated Z-0234-2022
- Terminated Z-0232-2022
7FR Percutaneous Thrombolytic Device (PTD) Inner lumen (orange) detached from the device basket result in embolization and retrieval with an additional intervention and consequent delay of therapy.
-
- Terminated Z-1502-2009
Leakage: Material not lipid resistant, may crack and/or leak. Use of this device with lipid containing solutions or 70% isopropyl alcohol may cause leakage or air embolism.
-
- Terminated Z-0842-05
The orange Pebax lumen located within the catheter basket may separate from the catheter basket sleeve permitting the lumen to become dislodged during use.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- QEW
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.5150
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent