Skip to content

Arrowadvantages Pressure Injectable Peripherally Inscrted Central Catheter Kits

by Arrow International, Inc.

510(k) K113277

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK113277 Date21 Jun 2012ClearedStatusActive

Recalls

6 recalls · 1 ongoing · FDA

    • Ongoing Z-2378-2026 FDA Class I initiated 21 Apr 2026

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

    • Terminated Z-1254-2020 FDA Class II initiated 20 Dec 2019

    Product lidstock contains the incorrect expiration date for the product

    • Terminated Z-2023-2018 FDA Class II initiated 11 Apr 2018

    Product sterility may be compromised due to unsealed packaging.

    • Terminated Z-0602-2018 FDA Class II initiated 25 Jul 2017

    Finished good kits may contain incorrect components. Kits that should contain a 5 FR 2 lumen catheter may contain a 4 FR single lumen catheter and kits that should contain a 4 FR single lumen catheter may contain a 5 FR 2 lumen catheter. This may cause a delay in treatment while a replacement catheter is obtained.

    • Terminated Z-0182-2018 FDA Class II initiated 1 Jun 2017

    There is a possibility that the finished good kits may contain 5 Fr peel-away sheaths instead of the required 4 Fr peel-away sheaths.

    • Terminated Z-0207-2016 FDA Class II initiated 16 Oct 2015

    Possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter. Although the issue does not pose a significant risk to the patient, it may cause a delay of treatment when placing the catheter.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LJS
Device Class
Class II
Regulation
21 CFR Part 880.5970
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent

Other devices from Arrow International, Inc.