Artemis Hand Held Imaging System
by Quest Medical Imaging 510(k)
Active
510(k) K143474
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK143474 | Date30 Jun 2015 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-0232-2025
Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention
Specification
- 510(k) Number
- FDA Product Code
- OWN
- Device Class
- Class II
- Regulation
- 21 CFR Part 876.1500
- Advisory Committee
- Gastroenterology, Urology
- FDA Decision
- SESE Substantially Equivalent