Arterial Line Kit
by Argon Medical Corp. 510(k)
Active Open recall
510(k) K780871
Regulatory status
Recall ongoing. 2 ongoing FDA recalls name this device. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK780871 | Date14 Jun 1978 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-2708-2026
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
-
- Ongoing Z-0563-2022
Guidewire included with two lots of Arterial Line Kits would not fit through the needle of the same kit. This could potentially lead to a slight delay in the procedure and minor blood loss.
-
- Terminated Z-1937-2020
- Terminated Z-1936-2020
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.
-
- Terminated Z-1745-2010
- Terminated Z-1743-2010
- Terminated Z-1744-2010
- Terminated Z-1741-2010
- Terminated Z-1737-2010
- Terminated Z-1742-2010
- Terminated Z-1740-2010
There are 1-2 places on some of the multi-purpose kit blisters/trays in which there are holes and/or the material is so thin that holes are easily created. This would potentially cause the sterility of the products to be compromised.
Specification
- 510(k) Number
- FDA Product Code
- GDY
- Device Class
- Class I
- Regulation
- 21 CFR Part 878.4450
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent