# ARTIS icono ceiling

> Active: European Union and United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/artis-icono-ceiling-11328100>
- Manufacturer: [Siemens Healthcare GmbH](https://devicerts.com/manufacturers/siemens-healthcare-gmbh) (Germany)
- Device class: Class IIb (EU MDR); Class II (US FDA)
- FDA product code: IZI, JAA, OWB, JAK
- EMDN: Z11030102 STATIONARY ANGIOGRAPHY SYSTEMS FOR ANGIOGRAPHIC AND CARDIOLOGIC STUDIES
- GMDN: Stationary angiographic x-ray system, digital (37623)

## Registrations and authorizations

- **EUDAMED** · CE/MDR · European Union · Basic UDI-DI 0405686900146UY · UDI-DI 04056869295923 · registered 2026-04-27 · status: Active
- **GUDID** · FDA 510(k) · United States · K254173 · decision 2026-04-24 · decision SESE · status: Active · from GUDID
- **GUDID** · FDA 510(k) · United States · K241572 · decision 2024-10-22 · decision SESE · status: Active · from GUDID
- **GUDID** · GUDID listing · United States · UDI-DI 04056869295923 · listed 2022-10-01 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K221516 · decision 2022-07-19 · decision SESE · status: Active · from GUDID

## Recalls

- Z-0523-2024 · Class II · ongoing · initiated 2023-10-17: Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.
- Z-2648-2026 · Class II · ongoing · initiated 2026-06-08: Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
- Z-2641-2026 · Class II · ongoing · initiated 2026-05-20: Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
- Z-2642-2026 · Class II · ongoing · initiated 2026-05-20: Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
- Z-2643-2026 · Class II · ongoing · initiated 2026-05-20: Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
- Z-1814-2026 · Class II · ongoing · initiated 2026-03-12: During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
- Z-1653-2026 · Class II · ongoing · initiated 2026-03-10: During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
- Z-2205-2026 · Class II · ongoing · initiated 2026-03-09: During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
- Z-2206-2026 · Class II · ongoing · initiated 2026-03-09: During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
- Z-2529-2025 · Class II · ongoing · initiated 2025-08-12: Limited system movements after startup .
- Z-2020-2025 · Class II · ongoing · initiated 2025-05-15: In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- Z-2021-2025 · Class II · ongoing · initiated 2025-05-15: In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- Z-2022-2025 · Class II · ongoing · initiated 2025-05-15: In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- Z-2241-2024 · Class II · ongoing · initiated 2024-06-26: A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.

## Identity

- Generic name: Stationary angiographic x-ray system, digital [FDA GUDID (AccessGUDID)]
- Device name: MDR\_01\_IIb\_FOR\_ACDV\_ARTIS icono/pheno [EUDAMED]
- Model / reference: MDR\_01\_IIb\_FOR\_ACDV\_ARTIS icono/pheno [EUDAMED]; 11328100 [EUDAMED, FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 04056869295923 [EUDAMED, FDA GUDID (AccessGUDID)]
- Basic UDI-DI: 0405686900146UY [EUDAMED]
- 510(k) Number: K221516 [FDA GUDID (AccessGUDID)]; K241572 [FDA GUDID (AccessGUDID)]; K254173 [FDA GUDID (AccessGUDID)]
- FDA product code: IZI [FDA GUDID (AccessGUDID)]; JAA [FDA GUDID (AccessGUDID)]; OWB [FDA GUDID (AccessGUDID)]; JAK [FDA GUDID (AccessGUDID)]
- EMDN code: Z11030102 [EUDAMED]
- GMDN term: Stationary angiographic x-ray system, digital (37623) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class IIb [EUDAMED]; Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 892.1600 [FDA GUDID (AccessGUDID)]; 892.1650 [FDA GUDID (AccessGUDID)]; 892.1750 [FDA GUDID (AccessGUDID)]
- EU registration type: Standard device [EUDAMED]
- EU market of registration: Germany [EUDAMED]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Measuring function: Yes [EUDAMED]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Sterilise before use: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reusable: No [EUDAMED]
- Contains latex: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reprocessed single-use device: No [EUDAMED]
- Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction): No [EUDAMED]
- Incorporates a substance derived from human blood or plasma: No [EUDAMED]
- Contains human tissues or cells: No [EUDAMED]
- Contains animal tissues or cells: No [EUDAMED]
- Contains endocrine-disrupting substances: No [EUDAMED]
- Administering medicine: No [EUDAMED]
- Incorporates a medicinal substance: No [EUDAMED]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: ARTIS icono ceiling, Siemens Healthcare GmbH. deviCERTS medical device catalogue. https://devicerts.com/devices/artis-icono-ceiling-11328100 (data as of 2026-11-29).
- Data sources: [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-25
- Statuses computed as of: 2026-11-29
- JSON: <https://devicerts.com/devices/artis-icono-ceiling-11328100.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/artis-icono-ceiling-11328100
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.