ARTIS icono (Ceiling Configuration)
by Siemens Medical Solutions USA, Inc.
510(k) K221516
ARTIS icono (Ceiling Configuration) is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K221516 on 19 Jul 2022, active as of 22 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK221516 | Date19 Jul 2022 | StatusActive |
Official records
- FDA 510(k) database K221516
Recalls
-
- Ongoing Z-2648-2026
Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
-
- Ongoing Z-2241-2024
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
-
- Ongoing Z-0523-2024
Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.
Timeline
- Cleared FDA 510(k) K221516
- Recall initiated FDA recalls Z-0523-2024
- Recall initiated FDA recalls Z-2241-2024
- Recall initiated FDA recalls Z-2648-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- OWB
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent