ARTIS Icono
by Siemens Medical Solutions USA, Inc. 2 variants
510(k) K193326
ARTIS Icono is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K193326 on 16 Apr 2020, active as of 22 Oct 2026. 12 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
6 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK193326 | Date16 Apr 2020 | StatusActive |
Official records
- FDA 510(k) database K193326
Recalls
-
- Ongoing Z-1342-2022
During system tests, an increased wearing of the Image acquisition system fans has been observed. This could lead to an electrical failure which may cause malfunction of the image system during regular system operation.
-
- Ongoing Z-1245-2022
- Ongoing Z-1246-2022
if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm
-
- Ongoing Z-0985-2022
- Ongoing Z-0986-2022
Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed
-
- Ongoing Z-1339-2022
After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient orientation/position may be misinterpreted and result in inappropriate treatment, even if the incorrect visualization is obvious.
-
- Terminated Z-0516-2022
- Terminated Z-0517-2022
Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option
-
- Terminated Z-1497-2021
Siemens has become aware of a potential issue with software version VE20C. Planned procedures may have to be terminated and performed on an alternative x-ray system.
-
- Terminated Z-1288-2021
Siemens has become aware of a potential issue in which the imaging system UPS sporadically forces a shutdown of the imaging system PC without a true power supply problem.
-
- Terminated Z-1144-2021
Misleading error messages and a gap in the Operator Manual which affects ARTIS Icono biplane or ARTIS Icono floor systems with software version VE20B. Potential issues include System error management, Erroneous error messages, Zoom/Pan Function, Grid Indication, and Coolant Level. May cause procedures to be terminated and performed on an alternative x-ray system
-
- Terminated Z-0400-2021
There is a software problem which affects the DSA Roadmap application on ARTIS Icono and ARTIS Pheno systems with software version VE20B. In rare cases, when the DSA acquisition has been started while the C-Arm or table is moving, there may be a shift between the image acquired at the start position and the image (i.e. DSA vessel map) taken at the reference position. This may lead to a DSA vessel map being overlaid to a subtracted fluoro image at a position that is not accurate.. However, if the displacement is slight or at an unfavorable plane, the user might rely on incorrect visualization of the catheter relative to the vessel map. This could imply danger to the patient.
-
- Terminated Z-2257-2020
Using Quantification Application (QVA/QCA) on DSA images may lead to a failure in vessel detection or incorrect quantification of vessel detection and result in an incorrect diagnosis and inappropriate treatment of the patient
Timeline
- Recall initiated FDA recalls Z-1288-2021
- Recall initiated FDA recalls Z-1497-2021
- Recall initiated FDA recalls Z-0516-2022
- Recall initiated FDA recalls Z-0517-2022
- Recall initiated FDA recalls Z-0985-2022
- Recall initiated FDA recalls Z-0986-2022
- Recall initiated FDA recalls Z-1339-2022
- Recall initiated FDA recalls Z-1245-2022
- Recall initiated FDA recalls Z-1246-2022
- Recall initiated FDA recalls Z-1342-2022
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- OWB
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent