Artis One
by Siemens Medical Solutions USA, Inc.
510(k) K133580
Artis One is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K133580 on 28 Apr 2014, active as of 22 Oct 2026. 7 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK133580 | Date28 Apr 2014 | StatusActive |
Official records
- FDA 510(k) database K133580
Recalls
-
- Ongoing Z-0922-2025
A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.
-
- Ongoing Z-1469-2023
Examination room monitor connected with the hybrid cable may have a display issue and no image function is available
-
- Terminated Z-0298-2021
Potential malfunction in which the examination room monitor may become intermittently dysfunctional (e.g., no display/flicking/wrong resolution).
-
- Terminated Z-1148-2018
Certain patient table systems may have a missing or incorrectly installed snap ring that could cause the bolt to slide out of its base and the tabletop carriage to slide to the floor.
-
- Terminated Z-0614-2018
The possibility exists for the monitor display of Artis one systems with a DP Transceiver below revision 03 to experience blinking and/or a blue/black screen
-
- Terminated Z-1282-2016
After importing segmentation results of the left atrium created on the Artis One system, the possibility exists for the results to be mirrored to an electro-anatomical 3D mapping system, CARTO from Biosense Webster Inc.
-
- Terminated Z-1119-2015
The possibility exists that the monitor may fail and requires a power circle (shutdown and then power on) to resume operation. The problem is not systematic; but sporadic on single units.
Timeline
- Cleared FDA 510(k) K133580
- Recall initiated FDA recalls Z-1119-2015
- Recall initiated FDA recalls Z-1282-2016
- Recall initiated FDA recalls Z-0614-2018
- Recall initiated FDA recalls Z-1148-2018
- Recall initiated FDA recalls Z-0298-2021
- Recall initiated FDA recalls Z-1469-2023
- Recall initiated FDA recalls Z-0922-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- OWB
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent