Aspire Hd Full-Field Digital Mammography System

by Fujifilm Medical Systems U.S.A, Inc. 510(k)

Active United States

510(k) K110729

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK110729 Date1 Sep 2011ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

3 recalls · FDA

    • Terminated Z-0659-2018 FDA Class II initiated 4 Aug 2017

    FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR1000AWS, FDR-2000AWS, FDR-3000AWS, and CRIR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.

    • Terminated Z-2690-2016 FDA Class II initiated 9 Jun 2016

    If an exposure is aborted during a stereo biopsy workflow, an error may occur which would prevent the system from continuing to operate properly.

    • Terminated Z-1585-2015 FDA Class II initiated 3 Feb 2015

    User Manual lacks a description for handling some errors even though the system can be recovered from such errors

Specification

510(k)

510(k) Number
FDA Product Code
MUE
Device Class
Class II
Regulation
21 CFR Part 892.1715
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent