> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/astral-100-150-cbk>
- Manufacturer: [Resmed, Ltd.](https://devicerts.com/manufacturers/resmed-corp) (United States)
- Device class: Class II (US FDA)
- FDA product code: CBK

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K172875 · decision 2018-04-26 · decision SESE · status: Active

## Recalls

- Z-2735-2026 · Class I · ongoing · initiated 2026-06-25: Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
- Z-0111-2024 · Class I · ongoing · initiated 2023-09-13: If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered by a supercapacitor, which degrades over time, which may cause TPF alarm to sound for less than 2 minutes or not at all

## Identifiers

- 510(k) Number: K172875 [FDA 510(k)]
- FDA product code: CBK [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 868.5895 [FDA 510(k)]
- Advisory committee: Anesthesiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Astral 100/150, Resmed, Ltd. deviCERTS medical device catalogue. https://devicerts.com/devices/astral-100-150-cbk (data as of 2026-11-19).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-19
- JSON: <https://devicerts.com/devices/astral-100-150-cbk.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/astral-100-150-cbk