Atellica CH 930 Analyzer
by Siemens Healthcare Diagnostics Inc. EUDAMEDGUDID
Active
Ref 11768553Model IVDR_67_TTN_DUB-C_A_INST_930 Analyzer
Regulatory status
Recall ongoing. 5 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDR EUDAMED | Number0405686901510V6 | Date6 Jul 2026 | StatusActive |
| United States | GUDID listing GUDID | Number00630414651637 | Date23 Oct 2025 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
- European Union
- United States
Certificates
No notified-body certificate is listed. Devices of this class are self-declared by the manufacturer.
Recalls
-
- Ongoing Z-2027-2025
Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
-
- Ongoing Z-1003-2024
Potential for a negative bias with quality control (QC) and patient sample results when using the Atellica CH Immunoglobulin M_2 (IgM_2) reagent.
-
- Ongoing Z-0449-2022
(1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay Correlation factor; (2) On Board Stability (OBS) Not Updating with Manual Change for Open Channel Assays may continue to use the reagent past its expiration date leading to potential erroneous patient sample results
-
- Ongoing Z-2026-2021
Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject to high pressure or high wear and may impact the result accuracy of any assay.
-
- Ongoing Z-1870-2021
Photometer Lamp May Reach Saturation Without Flagging Results, may generate erroneously elevated or depressed patient results
-
- Terminated Z-0861-2021
Multiple issues in Software v1.23.2 and lower-may lead to the reporting of discrepant patient results
-
- Terminated Z-1797-2020
Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test definition, and Auto-Rerun or Auto-Dilution results may generate a result of Zero (0) for quantitative assays or Negative for qualitative assays if a well of the reagent pack is not calibrated.
-
- Terminated Z-0862-2020
On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods.
-
- Terminated Z-0581-2020
Potential for calibration error resulting in QC failures after the calibration run with QC and patient results being 5 times higher.
-
- Terminated Z-0148-2020
Atellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below
-
- Terminated Z-0782-2019
Multiple Issues Identified in Atellica Solution System Software in V 1.14.2 and lower in: Atellica IM 1300 Analyzer Atellica IM 1600 Analyzer;Atellica CH 930 Analyzer; Atellica Sample Handler Prime
-
- Terminated Z-0194-2019
Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific Reagent kits are loaded on the analyzer at the same time. 2. Sample barcodes may be incorrectly read 3. The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not run CEA dilutions at 50x and 100x. The Analyzer may post a Sample Integrity Error when 50x or 100x dilutions are ordered for Carcinoembryonic Antigen (CEA). Dilution factors lower than 50x and neat samples are not affected by this behavior. 4. The Atellica CH 930 Analyzer may continue to display the default reporting units after the reporting units have been changed. 5. The Atellica CH 930 Analyzer may not display results for sample if certain assays are configured to auto-dilute results above or below the assay measuring interval.
Specification
- Trade Name
- Atellica CH 930 Analyzer EUDAMEDAtellica CH930 GUDID
- Generic Name
- Multiple clinical chemistry analyser IVD, laboratory
- Device Name
- IVDR_67_TTN_DUB-C_A_INST_930 Analyzer
- Model / Ref / Catalog
- IVDR_67_TTN_DUB-C_A_INST_930 Analyzer, 11768553, 11067000
- Primary DI / UDI-DI
- 00630414651637
- Basic UDI-DI
- 0405686901510V6
- 510(k) Number
- FDA Product Code
- JJE
- EMDN Code
- W0201010107 CHEMISTRY ANALYSERS - MODERATE TO HIGH ROUTINE (with ISE mode test/h > 800 and <= 1.200)
- GMDN Term
- Multiple clinical chemistry analyser IVD, laboratory
- Device Class
- Class A EUDAMEDClass I GUDID
- Regulation Number
- 862.2160
- Market of Registration
- Belgium
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class A | Class I |
Manufacturer details
- EUDAMED SRN
- US-MF-000016560
- Authorised representative
- Siemens Healthcare Diagnostics Manufacturing Limited IE-AR-000006763
Other devices from Siemens Healthcare Diagnostics Inc.
Properties
Other properties (21)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Human Tissues: No
- Animal Tissues: No
- Microbial Substances: No
- Reagent: No
- Self Testing: No
- Near-Patient Testing: No
- Companion Diagnostics: No
- New Device: No
- Prescription Use: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No