FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK961771 Date25 Sep 1996ClearedStatusActive

Recalls

1 recall · 1 ongoing · FDA

    • Ongoing Z-1411-2021 FDA Class II initiated 5 Mar 2021

    Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
GEI
Device Class
Class II
Regulation
21 CFR Part 878.4400
Advisory Committee
General, Plastic Surgery
FDA Decision
SESE Substantially Equivalent

Overview

AI-generated summary, 1 Oct 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The Auto Suture Surgiwand is a surgical automatic clip applier, classified under product code GEI. It is designed for the ligation and transection of tissue, facilitating precise and controlled closure of blood vessels and other anatomical structures during surgical procedures. This device is intended for use in general and plastic surgery applications, providing a reliable method for securing surgical sites.

The Surgiwand is supplied as a reusable surgical instrument and is intended for use in sterile operating environments. It is not labeled as MRI-safe, and its stainless steel construction requires proper cleaning, sterilization, and maintenance between uses. The device holds an FDA 510(k) clearance (K961771) under the Traditional pathway, with a clearance date of September 25, 1996. No additional regulatory markings or sterility assurances are specified in the provided data.

Other devices from United States Surgical, A Division of Tyco Healthc

See all devices from United States Surgical, A Division of Tyco Healthc →

Questions

AI-generated answers, 1 Oct 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What is the Auto Suture Surgiwand used for?

    The Auto Suture Surgiwand is designed specifically for the ligation and transection of tissue, enabling surgeons to securely close blood vessels and other anatomical structures during surgical procedures. Its precise and controlled application helps facilitate reliable surgical site closure in general and plastic surgery settings.

  • Is the Auto Suture Surgiwand reusable or single-use?

    The Auto Suture Surgiwand is supplied as a reusable surgical instrument. Due to its stainless steel construction, it requires thorough cleaning, sterilization, and maintenance between uses to ensure proper functionality and infection control in sterile operating environments.

  • What materials is the Auto Suture Surgiwand made from?

    The Auto Suture Surgiwand is constructed from stainless steel. This material choice ensures durability and compatibility with standard sterilization processes, though it also means the device is not labeled as MRI-safe.

  • How is the sterility of the Auto Suture Surgiwand maintained?

    The device itself is not labeled with specific sterility assurances in the provided data. However, as a reusable instrument, its sterility is maintained through proper cleaning, sterilization, and maintenance protocols between uses in a sterile operating environment.

  • What regulatory pathway was followed for the Auto Suture Surgiwand?

    The Auto Suture Surgiwand received FDA clearance under the Traditional 510(k) pathway, classified as substantially equivalent to a predicate device. The clearance (K961771) was granted on September 25, 1996, and falls under product code GEI for general and plastic surgery applications.