Avea Ventilator
by Care Fusion
510(k) K103211
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK103211 | Date12 May 2011 | StatusActive |
Recalls
-
- Terminated Z-1894-2016
- Terminated Z-1888-2016
- Terminated Z-1892-2016
- Terminated Z-1897-2016
- Terminated Z-1895-2016
- Terminated Z-1896-2016
- Terminated Z-1898-2016
- Terminated Z-1899-2016
- Terminated Z-1889-2016
- Terminated Z-1890-2016
- Terminated Z-1893-2016
- Terminated Z-1891-2016
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
-
- Terminated Z-1609-2015
A pressure transducer failure can develop, activating a false Extended High Ppeak or Circuit Occlusion Alarm. The Safety Valve will open to ambient air and the unit will stop ventilating, allowing spontaneous breathing patients to breathe. The malfunction may delay initiation or cease ventilation. Nonbreathing patients will need manual ventilation or to be connected to another ventilator.
-
- Terminated Z-2273-2013
CareFusion has identified a potential risk associated with AVEA ventilators when used at higher altitudes. A proactive complaint review identified an error in the barometric pressure compensation. CareFusion is voluntarily initiating a field correction of the affected devices to correct the error.
-
- Terminated Z-0510-2012
The recall was initiated because Carefusion has identified a potential risk associated with certain AVEA ventilators and affected replacement parts manufactured between March 1, 2009 and June 30, 2011. Carefusion is voluntarily initiating a field correction of the affected devices to preclude the possibility of this risk.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CBK
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent