Avl 9180 Electrolyte Analyzer
by Avl Scientific Corp. 510(k)
Active
510(k) K961458
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK961458 | Date12 Jun 1996 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1098-2021
May display Calcium results on the screen with wrong arrow direction which may lead to misinterpretation of results and incorrect medical decision. If the unit for iCa++ is set to mg/dL (configured as Service Code MGL) and a measured iCa++ value is lower than the normal range, an upward arrow is shown on the display instead of a downward arrow. The display is set per default to mmol/L; therefore, the arrow indicating an alarm reflects normal ranges of iCa++ in mmol/L. If mg/dL is chosen, the arrow warnings on screen still reflect the ranges from mmol/L, rather than mg/dL. The numeric result is displayed correctly.
Specification
- 510(k) Number
- FDA Product Code
- JFP
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1145
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent