Avon 6430-0-040
Avon 6430-0-040 is a device of GMDN type polyethylene patella prosthesis (FDA Class II) made by Howmedica Osteonics Corp., cleared under FDA 510(k) K052917 on 27 Dec 2005 (2 earlier submissions) and listed in FDA GUDID since 24 Sep 2015 (UDI-DI 07613327049329), active as of 23 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
7 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number07613327049329 | Date24 Sep 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK052917 | Date27 Dec 2005 | StatusActive |
Official records
- AccessGUDID record 07613327049329
- FDA 510(k) database K052917
- FDA 510(k) database K051948
- FDA 510(k) database K020841
From the FDA record
- Device description
- Femoral Joint Patella Component
- MRI safety
- MR Conditional
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K052917 decided 27 Dec 2005 Active
- FDA GUDID UDI-DI 07613327049329 listed 24 Sep 2015 Listed
Recalls
-
- Ongoing Z-3206-2026
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
- Ongoing Z-0895-2024
- Ongoing Z-0894-2024
- Ongoing Z-0896-2024
- Ongoing Z-0893-2024
- Ongoing Z-0897-2024
- Ongoing Z-0900-2024
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
- Terminated Z-0606-2012
Two National Joint Registries suggest that the EIUS Unicompartmental System is associated with a higher revision rate then the unicompartmental devices.
Timeline
- Cleared FDA 510(k) K052917
- Recall initiated FDA recalls Z-0606-2012
- Listed FDA GUDID 07613327049329
- Recall initiated FDA recalls Z-0895-2024
- Recall initiated FDA recalls Z-0894-2024
- Recall initiated FDA recalls Z-0896-2024
- Recall initiated FDA recalls Z-0893-2024
- Recall initiated FDA recalls Z-0897-2024
- Recall initiated FDA recalls Z-0900-2024
- Recall initiated FDA recalls Z-3206-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Avon
- Generic name
- Polyethylene patella prosthesis
- Model / Ref / Catalog
- 6430-0-040
- Primary DI / UDI-DI
- 07613327049329
- FDA product code
- HSX, HRY, KRR, NPJ
- GMDN term
- Polyethylene patella prosthesis
- Device class
- Class II
- Regulation number
- 888.3520, 888.3530, 888.3540, 888.3560
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Conditional
Similar devices in this category
Other devices filed under Polyethylene patella prosthesis.
- Patella Components 15-2521/40 by WALDEMAR LINK GmbH & Co. KG
- Patella Components 318-401/28 by WALDEMAR LINK GmbH & Co. KG
- APEX Knee System EC-40380 by OMNIlife science, Inc.
- OMNI Knee KC-40298 by OMNIlife science, Inc.
- PHYSICA Patellar Prosthesis 6595.50.035 by Limacorporate
- Arthrex® AR-504-PSA8 by Arthrex, Inc.
- MOTO PFJ System 02.15.E035 by Medacta International SA
- GMK Primary Total Knee System 02.07.0043IP by Medacta International SA
- Unity Total Knee System 112.018.46 by CORIN LTD
- PHYSICA Patellar Prosthesis 6595.50.038 by Limacorporate