Aw Server
by GE Medical Systems, LLC United States 510(k)
510(k) K081985
Product code LLZ
Regulatory status
Recall ongoing. 2 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK081985 | Date25 Sep 2008 | StatusActive |
Coverage
- Active
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-2553-2026
GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.
-
- Ongoing Z-0068-2026
Firm has identified a security vulnerability in AW Server products. If exploited, a malicious actor could compromise the confidentiality, integrity, and availability of patient data.
-
- Terminated Z-0270-2014
GE Healthcare has recently become aware of a potential safety issue due to the image orientation within GSI viewer on AW products. The GSI Viewer application running on AW/Volume Share or on the AW Server when making a batch and rotating the batch group clockwise 180 degrees (Inferior to Superior) will display images in the Superior orientation. The orientation annotation on the image is correc
-
- Terminated Z-1137-2013
GE Healthcare has recently become aware of a potential safety issue with respect to the Results Viewer in the AW Server. After starting the Results Viewer with the selected series of a patient, the pixel and certain annotation data from a different patient is displayed in one of the viewports. This occurs on an occasional basis, when multiple users are simultaneously accessing the Results Viewe
-
- Terminated Z-0486-2012
2 Unrelated Software Issues: An incorrect patient name maybe displayed when switching between two Volume Viewer tabs with different patients. Also, if a second patient exam is selected while the application is starting, the second patient's data will be displayed with the first patient's name display on the applications tab.
Specification
- 510(k) Number
- FDA Product Code
- LLZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.2050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Other devices from GE Medical Systems, LLC
Properties & raw data
Show properties, registry sources and raw records
Sources (1)
- FDA 510(k) K081985