# Axiom Luminos Drf

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/axiom-luminos-drf-owb>
- Manufacturer: [Siemens Medical Solutions USA, Inc.](https://devicerts.com/manufacturers/siemens-medical-solutions-usa-inc) (United States)
- Name in the registry: AXIOM LUMINOS DRF
- Device class: Class II (US FDA)
- FDA product code: OWB

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K062623 · decision 2007-08-22 · decision SESE · status: Active

## Recalls

- Z-1164-2023 · Class II · ongoing · initiated 2022-12-21: Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
- Z-1166-2023 · Class II · ongoing · initiated 2022-12-21: Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
- Z-1668-2015 · Class II · terminated · initiated 2015-04-13: Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
- Z-0994-2015 · Class II · terminated · initiated 2014-12-15 · matched through shared submission K062623, may concern other models: Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.
- Z-0003-2015 · Class II · terminated · initiated 2014-09-02 · matched through shared submission K062623, may concern other models: Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
- Z-2649-2014 · Class II · terminated · initiated 2014-08-12: It was determined that a potential malfunction may occur on the Luminos dRF system with SW VD10 or on Luminos dRF Max when the table side control is exposed to fluids. Fluid may infiltrate the tableside control panel and result in malfunction. In the very unlikely case of fluid infiltration a permanent short circuit within the control panel of the Luminos dRF / Luminos dRF Max will automatically d
- Z-2292-2014 · Class II · terminated · initiated 2014-07-07 · matched through shared submission K062623, may concern other models: It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in longer radiation exposure than required. This may also result in an overexposed image that is not of diagnostic quality. As a result the taken examination needs to be repeated. Thi
- Z-1711-2014 · Class II · terminated · initiated 2014-04-11: It was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) \>8 kV may trigger an unintended movement of the AXIOM Luminos dRFsystem which may lead to an emergency situation and pose danger to the patients, operating personnel or to the unit.
- Z-0140-2014 · Class II · terminated · initiated 2013-09-03 · matched through shared submission K062623, may concern other models: When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in radiation continuing longer than required resulting in an overexposed image that is not of diagnostic quality, prompting repetition of the examination.
- Z-0374-2014 · Class II · terminated · initiated 2013-09-03 · matched through shared submission K062623, may concern other models: A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed images that are not of diagnostic quality.
- Z-1845-2014 · Class II · terminated · initiated 2013-08-28 · matched through shared submission K062623, may concern other models: When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination.
- Z-1848-2013 · Class II · terminated · initiated 2013-05-07: Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may collide with the floor. Four consignees were notified of this recall by Update Instruction XP067/12/S -- all four units have been corrected.
- Z-0016-2014 · Class II · terminated · initiated 2013-05-06: Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10. During a RAD examination the possibility exists that radiation may continue longer than expected and may result in unnecessary dose applied during an acquisition. Additionally, it is not possible for the user to stop the radiation manually; ho
- Z-1383-2012 ·  · terminated · initiated 2012-02-16 · matched through shared submission K062623, may concern other models: During regular product monitoring, Siemens became aware of a potential issue with the AXIOM Luminos dRF. Firm became aware of a potential issue with torque wrenches - during a regular scheduled calibration of the wrench, it was discovered that the wrench was out of the specification. The wrench was over-torquing approximately 21% for low-end readings (50 ftlbs) and 9% for high-end readings (250
- Z-1159-2012 ·  · terminated · initiated 2012-02-15 · matched through shared submission K062623, may concern other models: Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images may display dark and not suitable for diagnostic. Due to this software error customers have to repeat x-ray exposure to receive diagnostic images.
- Z-1421-2011 ·  · terminated · initiated 2011-01-14 · matched through shared submission K062623, may concern other models: Firm became aware of a potential malfunction and possible hazard to patients in the event the tilt buttons on the OPTI Grip become stuck under the housing
- Z-0717-2011 ·  · terminated · initiated 2010-10-25 · matched through shared submission K062623, may concern other models: patient leg can be pinched
- Z-2222-2010 ·  · terminated · initiated 2010-07-13 · matched through shared submission K062623, may concern other models: Fluid may enter the multi-functional handle (Opti Grip) when using the Axiom Luminos TF or Sireskop SD system, causing unintended reaction or movement, increased radiation or useless images.
- Z-0327-2009 ·  · terminated · initiated 2008-09-19 · matched through shared submission K062623, may concern other models: Issue may occur where an intended movement of the joystick may unintentionally initiate an x-ray exposure.
- Z-0326-2009 ·  · terminated · initiated 2008-09-19 · matched through shared submission K062623, may concern other models: Issue may occur where an intended movement of the joystick may unintentionally initiate an x-ray exposure.

## Identifiers

- 510(k) Number: K062623 [FDA 510(k)]
- FDA product code: OWB [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.1650 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Axiom Luminos Drf, Siemens Medical Solutions USA, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/axiom-luminos-drf-owb (data as of 2026-11-16).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-16
- JSON: <https://devicerts.com/devices/axiom-luminos-drf-owb.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/axiom-luminos-drf-owb
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.