Axsym Gentamicin
510(k) K935376
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK935376 | Date21 Jun 1994 | StatusActive |
Recalls
-
- Terminated Z-0930-2008
Incorrect Results -- Through evaluation of stability, Abbott have found a decrease in the calibrator A to F span with these reagents lots, which may produce the following results: - Printed Error Code "SPAN LESS THAN MIN SPAN" - Controls out of range. When a control is out of range, patient results should not be reported.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LCD
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.3450
- Advisory Committee
- Clinical Toxicology
- FDA Decision
- SESE Substantially Equivalent