Axsym Rubella Igg Antibody Assay
by Abbott Laboratories 510(k)
Active
510(k) K954045
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK954045 | Date24 Sep 1996 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2175-2008
An increase in complaints was noted for AxSYM Rubella IgG assay calibration failures due to error codes related to elevated Calibrator A (or Master Calibrator 1) rates too high.
-
- Terminated Z-1025-2007
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
Specification
- 510(k) Number
- FDA Product Code
- LFX
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.3510
- Advisory Committee
- Microbiology
- FDA Decision
- SESE Substantially Equivalent