Axsym Total B-Hcg
by Abbott Laboratories 510(k)
Active
510(k) K935673
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK935673 | Date21 Jun 1994 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1070-2007
- Terminated Z-1067-2007
- Terminated Z-1068-2007
- Terminated Z-1069-2007
An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is magnified potentially yelding false positive results for pregnancy and or Dilution Inaccuracy.
-
- Terminated Z-0893-05
Abbott Diagnostics is investigation a recent increase in customer complaints for AxSYM calibration error code 1048 and controls out of range when using AxSYM Total B-hCG reagent lots. Testing has shown that protein aggregates are present in the specimen diluents of both IMx and the AxSYM Total B-hCG reagent kits. This is responsible for the calibration errors and shift in control values.
-
- Terminated Z-0006-05
Assay may give elevated patients results.
-
- Terminated Z-1436-04
Complaints have been received for AxSYM Total B-hCG Reagents lots 17202Q100, 17556Q100 and 17387Q100 failing calibrations.
Specification
- 510(k) Number
- FDA Product Code
- JHI
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1155
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent