Azurion 7 B12
by Philips Medical Systems Nederland B.V. Netherlands EUDAMEDGUDID
Ref 722067
Product code OWBEMDN Z11030103
Regulatory status
Recall ongoing. 41 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDD EUDAMED | NumberB-00884838085350 | Date26 Nov 2024 | StatusActive |
| United States | GUDID listing GUDID | Number00884838085350 | Date4 Dec 2017 | StatusIn distribution |
Coverage
- Expiring (legacy directive)
- Listed
- Not covered
- European Union
- United States
Recalls
-
- Ongoing Z-2931-2026
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
-
- Ongoing Z-2892-2026
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
- Ongoing Z-2225-2026
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
-
- Ongoing Z-2170-2026
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
-
- Ongoing Z-1344-2026
- Ongoing Z-1350-2026
- Ongoing Z-1346-2026
- Ongoing Z-1347-2026
- Ongoing Z-1348-2026
- Ongoing Z-1351-2026
- Ongoing Z-1349-2026
- Ongoing Z-1345-2026
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
- Ongoing Z-1091-2026
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
-
- Ongoing Z-2502-2025
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
-
- Ongoing Z-2134-2025
Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
-
- Ongoing Z-1969-2025
A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).
-
- Ongoing Z-1156-2025
- Ongoing Z-1142-2025
- Ongoing Z-1154-2025
- Ongoing Z-1157-2025
- Ongoing Z-1160-2025
- Ongoing Z-1153-2025
- Ongoing Z-1144-2025
- Ongoing Z-1158-2025
- Ongoing Z-1159-2025
Possibility of the patient falling from the table related to the mattress used on the patient table.
-
- Ongoing Z-1092-2025
- Ongoing Z-1093-2025
- Ongoing Z-1094-2025
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
- Ongoing Z-1145-2024
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
-
- Ongoing Z-1152-2024
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
-
- Ongoing Z-2286-2026
- Ongoing Z-2289-2026
- Ongoing Z-2291-2026
- Ongoing Z-2287-2026
- Ongoing Z-2290-2026
- Ongoing Z-2288-2026
- Ongoing Z-2284-2026
- Ongoing Z-2285-2026
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
- Ongoing Z-2403-2023
A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
-
- Ongoing Z-0874-2022
Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
-
- Ongoing Z-0886-2022
In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
-
- Terminated Z-1145-2020
Due to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction.
-
- Terminated Z-2205-2019
The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.
-
- Terminated Z-1997-2019
Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1). When performing Cine runs, the Azurion's automatic exposure control software may set technique factors that result in radiation emissions too low to obtain useful diagnostic images.
Specification
- Trade Name
- Azurion 7 B12 EUDAMEDAzurion GUDID
- Generic Name
- Stationary angiographic x-ray system, digital
- Device Name
- Azurion series
- Model / Ref / Catalog
- 722067, Azurion 7 B12
- Primary DI / UDI-DI
- 00884838085350
- Basic UDI-DI
- B-00884838085350
- FDA Product Code
- OWB, JAA
- EMDN Code
- Z11030103 Biplanar angiography systems
- GMDN Term
- Stationary angiographic x-ray system, digital
- Device Class
- Class IIb EUDAMEDClass II GUDID
- Regulation Number
- 892.1650
- Market of Registration
- Netherlands
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | 722067 | Azurion 7 B12 |
| Device Class | Class IIb | Class II |
Manufacturer details
- EUDAMED SRN
- NL-MF-000001489
Other devices from Philips Medical Systems Nederland B.V.
See all devices from Philips Medical Systems Nederland B.V. →
Properties & raw data
Show properties, registry sources and raw records
Other properties (21)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No