Azurion 7 B20
by Philips Medical Systems Nederland B.V. Netherlands EUDAMEDGUDID
Ref 722226
Product code OWBEMDN Z11030102
Regulatory status
Recall ongoing. 64 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number0884838BM328SX | Date15 Apr 2025 | StatusActive |
| United States | GUDID listing GUDID | Number00884838099272 | Date26 May 2020 | StatusIn distribution |
Coverage
- Active
- Listed
- Not covered
- European Union
- United States
Recalls
-
- Ongoing Z-2932-2026
Two software related issues affecting Philips Azurion R3.0 and R3.1 systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications.
-
- Ongoing Z-2931-2026
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
-
- Ongoing Z-2892-2026
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
- Ongoing Z-2225-2026
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
-
- Ongoing Z-2170-2026
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
-
- Ongoing Z-1348-2026
- Ongoing Z-1349-2026
- Ongoing Z-1346-2026
- Ongoing Z-1345-2026
- Ongoing Z-1350-2026
- Ongoing Z-1344-2026
- Ongoing Z-1351-2026
- Ongoing Z-1347-2026
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
- Ongoing Z-1091-2026
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
-
- Ongoing Z-1113-2026
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
-
- Ongoing Z-1079-2026
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
-
- Ongoing Z-0597-2026
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
-
- Ongoing Z-0719-2026
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
-
- Ongoing Z-2502-2025
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
-
- Ongoing Z-1969-2025
A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).
-
- Ongoing Z-1089-2025
- Ongoing Z-1091-2025
- Ongoing Z-1092-2025
- Ongoing Z-1088-2025
- Ongoing Z-1093-2025
- Ongoing Z-1087-2025
- Ongoing Z-1090-2025
- Ongoing Z-1094-2025
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
- Ongoing Z-1155-2025
- Ongoing Z-1159-2025
- Ongoing Z-1158-2025
- Ongoing Z-1154-2025
- Ongoing Z-1156-2025
- Ongoing Z-1153-2025
- Ongoing Z-1157-2025
- Ongoing Z-1160-2025
Possibility of the patient falling from the table related to the mattress used on the patient table.
-
- Ongoing Z-3165-2024
Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
-
- Ongoing Z-2953-2024
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.
-
- Ongoing Z-2534-2024
Potential Loss of Imaging (X-ray) Functionality and/or Longer Time to Perform Cold Restart
-
- Ongoing Z-1178-2024
- Ongoing Z-1177-2024
- Ongoing Z-1176-2024
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
-
- Ongoing Z-1152-2024
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
-
- Ongoing Z-1145-2024
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
-
- Ongoing Z-0835-2024
System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.
-
- Ongoing Z-0860-2024
Loss of connectivity between the FlexArm and the Table due to a software issue.
-
- Ongoing Z-0861-2024
Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
-
- Ongoing Z-2286-2026
- Ongoing Z-2289-2026
- Ongoing Z-2291-2026
- Ongoing Z-2287-2026
- Ongoing Z-2290-2026
- Ongoing Z-2288-2026
- Ongoing Z-2284-2026
- Ongoing Z-2285-2026
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
- Ongoing Z-0558-2024
- Ongoing Z-0559-2024
Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
-
- Ongoing Z-0229-2024
Philips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer) is missing.
-
- Ongoing Z-2587-2023
A damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.
-
- Ongoing Z-2403-2023
A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
-
- Ongoing Z-2485-2023
Loss of availability of the wireless foot switch during procedures.
-
- Ongoing Z-2284-2023
The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
-
- Ongoing Z-0476-2022
Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure
-
- Ongoing Z-0238-2022
3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software.
-
- Terminated Z-0737-2022
Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system, the following on-screen message is displayed to the user in the Philips StentBoost Live application: "Press the cine pedal until the acquisition stops". However, due to the incorrect configuration of the EPX database, the acquisition does not stop after 40 images, but instead it continues as long as the pedal is pressed.
-
- Completed Z-0734-2022
In the Azurion system, the user can add a new study to a patient by selecting the option "Add Study". The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
Specification
- Trade Name
- Azurion 7 B20 EUDAMEDAzurion GUDID
- Generic Name
- Stationary angiographic x-ray system, digital
- Device Name
- Azurion series
- Model / Ref / Catalog
- 722226, Azurion 7 B20
- Primary DI / UDI-DI
- 00884838099272
- Basic UDI-DI
- 0884838BM328SX
- Direct Marketing DI
- 00884838099272
- FDA Product Code
- OWB, JAA
- EMDN Code
- Z11030102 Stationary angiography systems for angiographic and cardiologic studies
- GMDN Term
- Stationary angiographic x-ray system, digital
- Device Class
- Class IIb EUDAMEDClass II GUDID
- Regulation Number
- 892.1650
- Market of Registration
- Netherlands
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | 722226 | Azurion 7 B20 |
| Device Class | Class IIb | Class II |
Manufacturer details
- EUDAMED SRN
- NL-MF-000001489
Other devices from Philips Medical Systems Nederland B.V.
See all devices from Philips Medical Systems Nederland B.V. →
Properties & raw data
Show properties, registry sources and raw records
Other properties (21)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No
Sources (2)
- EUDAMED b3f690f9-d4ef-4164-9cd1-1fed422455a1
- FDA UDI 5a003c14-83c2-4019-937a-a306c104a9e2