{
  "format": "devicerts.device.v1",
  "cite_as": "Azurion 7 M20, Philips Medical Systems Nederland B.V. deviCERTS medical device catalogue. https://devicerts.com/devices/azurion-7-m20-722079 (data as of 2026-11-22).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/azurion-7-m20-722079",
    "json_url": "https://devicerts.com/devices/azurion-7-m20-722079.json",
    "markdown_url": "https://devicerts.com/devices/azurion-7-m20-722079.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "EUDAMED",
        "url": "https://ec.europa.eu/tools/eudamed"
      },
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-22",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/azurion-7-m20-722079",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Azurion 7 M20",
    "registry_name": "Azurion 7 M20",
    "url": "https://devicerts.com/devices/azurion-7-m20-722079",
    "manufacturer": {
      "name": "Philips Medical Systems Nederland B.V.",
      "url": "https://devicerts.com/manufacturers/philips-medical-systems-nederland-bv",
      "country": "Netherlands",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "European Union and United States"
    },
    "classification": [
      {
        "class": "Class IIb",
        "system": "EU MDD"
      },
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Catalogue number",
        "values": [
          {
            "value": "722079",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "00884838085268",
            "registries": [
              "EUDAMED",
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Basic UDI-DI",
        "values": [
          {
            "value": "B-00884838085268",
            "registries": [
              "EUDAMED"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "OWB",
      "JAA"
    ],
    "emdn": [
      {
        "code": "Z11030102",
        "term": "STATIONARY ANGIOGRAPHY SYSTEMS FOR ANGIOGRAPHIC AND CARDIOLOGIC STUDIES"
      }
    ],
    "gmdn": [
      {
        "code": "37623",
        "term": "Stationary angiographic x-ray system, digital"
      }
    ],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "EUDAMED",
      "region": "European Union",
      "type": "CE/MDD",
      "number": "B-00884838085268",
      "number_label": "Basic UDI-DI",
      "device_udi_di": "00884838085268",
      "date": "2025-03-25",
      "date_kind": "registered",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K181830",
      "number_label": null,
      "device_udi_di": null,
      "date": "2018-08-02",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K172822",
      "number_label": null,
      "device_udi_di": null,
      "date": "2017-11-22",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "00884838085268",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2017-03-31",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K163715",
      "number_label": null,
      "device_udi_di": null,
      "date": "2017-01-26",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0294-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-10-06",
      "reason": "Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-0299-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-10-06",
      "reason": "Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-0293-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-10-06",
      "reason": "Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-0295-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-10-06",
      "reason": "Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-0297-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-10-06",
      "reason": "Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-0292-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-10-06",
      "reason": "Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-0298-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-10-06",
      "reason": "Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-0300-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-10-06",
      "reason": "Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-0296-2027",
      "classification": "",
      "status": "ongoing",
      "initiated": "2026-10-06",
      "reason": "Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2931-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-07-16",
      "reason": "Two software related issues affecting Azurion R2.2.10 Systems:  \n1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and \n2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-2892-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-06-17",
      "reason": "Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2225-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-04-28",
      "reason": "Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2170-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-03-31",
      "reason": "Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1345-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-01-26",
      "reason": "Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1349-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-01-26",
      "reason": "Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1346-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-01-26",
      "reason": "Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1350-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-01-26",
      "reason": "Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1344-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-01-26",
      "reason": "Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1351-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-01-26",
      "reason": "Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1348-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-01-26",
      "reason": "Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1347-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2026-01-26",
      "reason": "Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1091-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-12-12",
      "reason": "Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2503-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-08-01",
      "reason": "Philips Azurion and Allura systems configured\nwith a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-2502-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-08-01",
      "reason": "Philips Azurion and Allura systems configured\nwith a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2134-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-06-04",
      "reason": "Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-1969-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-05-28",
      "reason": "A software issue was identified in the internal communication process between the system\nsoftware and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1092-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1093-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1094-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1160-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Possibility of the patient falling from the table related to the mattress used on the patient table.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-1157-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Possibility of the patient falling from the table related to the mattress used on the patient table.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-1153-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Possibility of the patient falling from the table related to the mattress used on the patient table.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-1156-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Possibility of the patient falling from the table related to the mattress used on the patient table.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1142-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Possibility of the patient falling from the table related to the mattress used on the patient table.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-1154-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Possibility of the patient falling from the table related to the mattress used on the patient table.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-1158-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Possibility of the patient falling from the table related to the mattress used on the patient table.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-1144-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Possibility of the patient falling from the table related to the mattress used on the patient table.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1159-2025",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2025-01-13",
      "reason": "Possibility of the patient falling from the table related to the mattress used on the patient table.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-1145-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-01-02",
      "reason": "Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1152-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-01-02",
      "reason": "Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1151-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-01-02",
      "reason": "Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-2289-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-12-07",
      "reason": "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable.  The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2285-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-12-07",
      "reason": "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable.  The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2284-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-12-07",
      "reason": "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable.  The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2288-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-12-07",
      "reason": "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable.  The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2290-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-12-07",
      "reason": "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable.  The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2287-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-12-07",
      "reason": "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable.  The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2291-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-12-07",
      "reason": "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable.  The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2286-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-12-07",
      "reason": "During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable.  The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2587-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-08-15",
      "reason": "A damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.",
      "applies": "this device",
      "via": "K163715"
    },
    {
      "number": "Z-2403-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-07-19",
      "reason": "A foot switch pedal may get stuck in the active position when the user releases the pedal,\nresulting in the emission of unintended radiation",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1647-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-03-21",
      "reason": "Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available\nthere will be a sudden interruption of the procedure",
      "applies": "this device",
      "via": "K181830"
    },
    {
      "number": "Z-0874-2022",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2022-02-02",
      "reason": "Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.",
      "applies": "shared submission",
      "via": "K172822"
    },
    {
      "number": "Z-0471-2022",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2021-11-12",
      "reason": "(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available",
      "applies": "this device",
      "via": "K181830"
    },
    {
      "number": "Z-0213-2022",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2021-09-24",
      "reason": "When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.",
      "applies": "this device",
      "via": "K181830"
    },
    {
      "number": "Z-0886-2022",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-05-26",
      "reason": "In the Azurion system, the user can add a new study to a patient by selecting the option  Add\nStudy . The Add Study dialogue box is then displayed where the Patient Type is selected to\nperform the study. Due to a software defect, when the study is initiated by pressing  Start\nProcedure , the Patient Type changes inadvertently to a Patient Type different than the one\nselected as shown in the Table below. Patient type is one of the factors involved in the dose\ncontrol process. The incorrect patient type changes the technique factors to be used by the\nsystem without notification to the user.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1809-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-04-03",
      "reason": "When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.",
      "applies": "this device",
      "via": "K181830"
    },
    {
      "number": "Z-1801-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-03-05",
      "reason": "No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose..",
      "applies": "this device",
      "via": "K181830"
    },
    {
      "number": "Z-1145-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-12-23",
      "reason": "Due to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-2205-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-07-16",
      "reason": "The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.",
      "applies": "this device",
      "via": "K172822"
    },
    {
      "number": "Z-1130-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-03-12",
      "reason": "During manufacturing of an Azurion R2.0 system, it was found out that the cable set was not according to specification. This resulted in damage of the 24V power cable of the detector.",
      "applies": "this device",
      "via": "K181830"
    },
    {
      "number": "Z-1997-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-11-30",
      "reason": "Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1).\n\nWhen performing Cine runs, the Azurion's automatic exposure control software may set technique factors that result in radiation emissions too low to obtain useful diagnostic images.",
      "applies": "this device",
      "via": "K172822"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "Azurion 7 M20",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "Azurion",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Stationary angiographic x-ray system, digital",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Device name",
          "values": [
            {
              "value": "Azurion series",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Model / Ref / Catalog",
          "values": [
            {
              "value": "722079",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "Azurion 7 M20",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "00884838085268",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Basic UDI-DI",
          "values": [
            {
              "value": "B-00884838085268",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "OWB",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "JAA",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "EMDN code",
          "values": [
            {
              "value": "Z11030102",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Stationary angiographic x-ray system, digital",
              "code": "37623",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class IIb",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "892.1650",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "EU registration type",
          "values": [
            {
              "value": "Legacy device (under a directive)",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "EU market of registration",
          "values": [
            {
              "value": "Netherlands",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Measuring function",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Reusable",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Reprocessed single-use device",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction)",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Incorporates a substance derived from human blood or plasma",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains human tissues or cells",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains animal tissues or cells",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains endocrine-disrupting substances",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Administering medicine",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Incorporates a medicinal substance",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}
