Azurion R2.0
by Philips Medical Systems Nederland B.V.
510(k) K181830
Active
Valid until
No expiry
FDA recalls
36 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK181830 | Date2 Aug 2018 | StatusActive |
Recalls
-
- Ongoing Z-0293-2027
- Ongoing Z-0295-2027
- Ongoing Z-0297-2027
- Ongoing Z-0294-2027
- Ongoing Z-0292-2027
- Ongoing Z-0298-2027
- Ongoing Z-0300-2027
- Ongoing Z-0296-2027
- Ongoing Z-0299-2027
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
- Ongoing Z-2892-2026
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
- Ongoing Z-2225-2026
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
-
- Ongoing Z-2170-2026
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
-
- Ongoing Z-1349-2026
- Ongoing Z-1345-2026
- Ongoing Z-1350-2026
- Ongoing Z-1344-2026
- Ongoing Z-1351-2026
- Ongoing Z-1348-2026
- Ongoing Z-1347-2026
- Ongoing Z-1346-2026
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
- Ongoing Z-1091-2026
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
-
- Ongoing Z-2502-2025
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
-
- Ongoing Z-2134-2025
Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
-
- Ongoing Z-1969-2025
A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).
-
- Ongoing Z-1145-2024
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
-
- Ongoing Z-2285-2026
- Ongoing Z-2286-2026
- Ongoing Z-2289-2026
- Ongoing Z-2291-2026
- Ongoing Z-2287-2026
- Ongoing Z-2290-2026
- Ongoing Z-2288-2026
- Ongoing Z-2284-2026
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
- Ongoing Z-2403-2023
A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
-
- Ongoing Z-1647-2023
Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure
-
- Ongoing Z-0874-2022
Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
-
- Terminated Z-0471-2022
(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available
-
- Terminated Z-0213-2022
When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.
-
- Terminated Z-1809-2020
When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.
-
- Terminated Z-1801-2020
No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose..
-
- Terminated Z-1145-2020
Due to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction.
-
- Terminated Z-1130-2019
During manufacturing of an Azurion R2.0 system, it was found out that the cable set was not according to specification. This resulted in damage of the 24V power cable of the detector.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent