Balanced Knee System 161-1406A
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00822409030248 | Date23 Oct 2015 | StatusListed |
| United States | FDA 510(k)SESE | NumberK994370 | Date24 Mar 2000 | StatusActive |
Official records
- AccessGUDID record 00822409030248
- FDA 510(k) database K994370
Recalls
-
- Ongoing Z-1903-2020
- Ongoing Z-1904-2020
The locking mechanism of the tibial tray exhibited a manufacturing defect in which a uniform ridge was cut along the inner wall of the "T" and "Y" junctions, which may cause resistance when the Tibial Insert was being assembled.
-
- Terminated Z-2255-2014
- Terminated Z-2254-2014
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.
-
- Terminated Z-1610-2014
A femoral component containing pegs was found in a box for the pegless version.
-
- Terminated Z-1993-2009
Erroneous lot number on 3 of 28 units of Balanced Knee System. Three units from Lot Number 0055245 were mixed in with the recalled lot (0054891).
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Balanced Knee System
- Generic name
- Uncoated knee femur prosthesis, metallic
- Model / reference
- 161-1406A
- Primary DI / UDI-DI
- 00822409030248
- 510(k) Number
- FDA product code
- JWH
- GMDN term
- Uncoated knee femur prosthesis, metallic
- Device class
- Class II
- Regulation number
- 888.3560
- GUDID distribution status
- In Commercial Distribution
Other devices from Ortho Development Corp.
Properties
Other properties (9)
- Sterilise before use: No
- Contains latex: No
- Over-the-counter: No
- Convenience kit: No
- Combination product: No
- HCT/P: No
- Premarket exempt: No
- Serial number: No
- Manufacturing date: No