Skip to content

Balloon Applicator MAJ-675

by Olympus Medical Systems Corporation

Ref N5364300

Balloon Applicator MAJ-675 is a surgical instrument assist arm system (EU MDR Class I; FDA Class II) made by Olympus Medical Systems Corporation, registered in EUDAMED on 23 Oct 2026 (Basic UDI-DI 495317000041B5), cleared under FDA 510(k) K093395 on 17 Jun 2010 and listed in FDA GUDID since 21 Sep 2016 (UDI-DI 04953170356001), active as of 7 Nov 2026. 4 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE MDR 510(k)

Valid until

No date given

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

510(k) K093395

FDA submission, from GUDID

Trust

Unverified

Source EUDAMED, GUDID, synced 7 Nov 2026.

Regulatory status

as of registry sync 7 Nov 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number495317000041B5 Basic UDI-DI 04953170356001UDI-DIDate23 Oct 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number04953170356001 UDI-DI Date21 Sep 2016ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE GUDID from GUDID Substantially Equivalent · Traditional NumberK093395 Date17 Jun 2010DecidedStatusActive

Official records

From the FDA record

Device description
"MAJ-675 Balloon applicator for GF-UC30P"
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile; sterilise before use

From the GUDID record (AccessGUDID), as read on 7 Nov 2026, as the labeller entered it.

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI 495317000041B5 registered 23 Oct 2026 Active
  • FDA 510(k) K093395 decided 17 Jun 2010 Active
  • FDA GUDID UDI-DI 04953170356001 listed 21 Sep 2016 Listed

Recalls

4 recalls · 1 ongoing · FDA

    • Ongoing Z-0333-2024 FDA Class II initiated 25 Sep 2023

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    • Terminated Z-0500-2021 FDA Class II initiated 16 Oct 2020

    A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.

    • Terminated Z-1007-2019 FDA Class II initiated 17 Dec 2018

    The adhesive used in the repair of the endoscope was incorrectly prepared. It is unknown what affect the nonconforming adhesive mixture would have on the durability or effectiveness of the gluing operation for the endoscope repair.

    • Terminated Z-2818-2015 FDA Class II initiated 21 Aug 2015

    Insufficient information for specific cleaning accessories used for the cleaning and reprocessing of the GF-UCT180 ultrasound endoscope

Timeline

7 dated events

  1. Cleared FDA 510(k) K093395
  2. Recall initiated FDA recalls Z-2818-2015
  3. Listed FDA GUDID 04953170356001
  4. Recall initiated FDA recalls Z-1007-2019
  5. Recall initiated FDA recalls Z-0500-2021
  6. Recall initiated FDA recalls Z-0333-2024
  7. Registered EUDAMED 495317000041B5

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States FDA 510(k) Active

Specification

EUDAMED · GUDID

Trade name
Balloon Applicator Maj-675 EUDAMED
Balloon Applicator GUDID
Generic name
Surgical instrument assist arm system, manually-adjusted
Device name
Balloon
Model / reference
N5364300 EUDAMED
MAJ-675 GUDID
Primary DI / UDI-DI
04953170356001
Basic UDI-DI
495317000041B5
FDA product code
IYO, IYN, NWB, ITX, ODG
EMDN code
G0380 Digestive endoscopy devices - accessories
GMDN term
Surgical instrument assist arm system, manually-adjusted
Device class
Class I EUDAMED
Class II GUDID
Regulation number
892.1560, 892.1550, 876.1500, 892.1570
EU registration type
Standard device
EU market of registration
Germany
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 3 fields differ

FieldEUDAMEDGUDID
Model / referenceEUDAMED
N5364300
GUDID
MAJ-675
Device classEUDAMED
Class I
GUDID
Class II
Sterilise before useEUDAMED
No
GUDID
Yes

Registry differences

EUDAMED and FDA GUDID describe the same product (same UDI-DI) and state different values:

  • Sterilise before use UDI-DI 04953170356001: EUDAMED states No, FDA GUDID states Yes.

Device class is stated on each registry's own scale: Class I in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Similar devices in this category

Other devices filed under DIGESTIVE ENDOSCOPY DEVICES - ACCESSORIES.

Other devices from Olympus Medical Systems Corporation

See all devices from Olympus Medical Systems Corporation →

Properties

  • Sterilise before use NoEUDAMEDYesGUDID
  • Prescription use
  • Lot / batch number
  • Manufacturing date
Other properties (21, all answered No)