BALLOON3 Maj-233
by Olympus Medical Systems Corporation
Ref N5369331
BALLOON3 Maj-233 (N5369331) is an ultrasound endoscope balloon (EU MDD Class I; FDA Class II) made by Olympus Medical Systems Corporation, registered in EUDAMED on 7 Oct 2026 (Basic UDI-DI B-04953170463914), cleared under FDA 510(k) K051541 on 1 Jul 2005 and listed in FDA GUDID since 25 Feb 2025 (UDI-DI 04953170463914), active as of 24 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No date given
FDA recalls
1 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDD EUDAMED | NumberB-04953170463914 04953170463914 | Date7 Oct 2026 | StatusActive |
| United States | GUDID listing GUDID | Number04953170463914 | Date25 Feb 2025 | StatusListed |
| United States | FDA 510(k)SESE GUDID | NumberK051541 | Date1 Jul 2005 | StatusActive |
Official records
- EUDAMED public site B-04953170463914
- AccessGUDID record 04953170463914
- FDA 510(k) database K051541
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- Labelled as containing natural rubber latex
- Sterility
- Supplied sterile
Across registries
- EUDAMED Basic UDI-DI B-04953170463914 registered 7 Oct 2026 Active
- FDA 510(k) K051541 decided 1 Jul 2005 Active
- FDA GUDID UDI-DI 04953170463914 listed 25 Feb 2025 Listed
Recalls
-
- Ongoing Z-0333-2024
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
- Terminated Z-0507-2021
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
Timeline
- Cleared FDA 510(k) K051541
- Recall initiated FDA recalls Z-0507-2021
- Recall initiated FDA recalls Z-0333-2024
- Listed FDA GUDID 04953170463914
- Registered EUDAMED B-04953170463914
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Trade name
- BALLOON3 Maj-233 EUDAMEDBALLOON3 GUDID
- Generic name
- Ultrasound endoscope balloon, Hevea-latex
- Device name
- Balloon (For Ultrasonic Endoscope)
- Model / Ref / Catalog
- N5369331, MAJ-233
- Primary DI / UDI-DI
- 04953170463914
- Basic UDI-DI
- B-04953170463914
- FDA product code
- FDF, ITX
- EMDN code
- G030899 Instruments for gastro-intestinal endoscopy and eco-endoscopy, single-use - other
- GMDN term
- Ultrasound endoscope balloon, Hevea-latex
- Device class
- Class I EUDAMEDClass II GUDID
- Regulation number
- 876.1500, 892.1570
- EU registration type
- Legacy device (under a directive)
- EU market of registration
- Germany
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / reference | N5369331 | MAJ-233 |
| Device class | Class I | Class II |
Similar devices in this category
Other devices filed under INSTRUMENTS FOR GASTRO-INTESTINAL ENDOSCOPY AND ECO-ENDOSCOPY, SINGLE-USE - OTHER.
- RevoWave PX RWHJ-3545SJ by PIOLAX MEDICAL DEVICES, INC.
- OTSC Proctology Anchor 38 cm 200.61 by Ovesco Endoscopy AG
- Disposable Guide wires FST-GW-C1/C2 by FORESIGHT TECHNOLOGY CO., LTD.
- Air Water Channel Cleaning Adapter GAR105 by GA HEALTH COMPANY LIMITED
- Duette G35129 DT-6-5F by Cook Ireland, Ltd.
- Air/Water Valve GAR004B by GA HEALTH COMPANY LIMITED
- RevoWave PX RWUS-2555I by PIOLAX MEDICAL DEVICES, INC.
- Disposable Non-vascular Guidewires ENDO-GW-88-300-A1 by Jiangsu Power-endo Medical Technology Co., Ltd
- Disposable Biopsy Valves MD-A-BV02 by ZHEJIANG CHUANGXIANG MEDICAL TECHNOLOGY CO., LTD.
- Disposable Non-vascular Guidewires ENDO-GW-88-260-A1 by Jiangsu Power-endo Medical Technology Co., Ltd