Bard Infusor Pump
510(k) K883577
Bard Infusor Pump is a medical device (FDA Class II; FDA product code FRN, Pump, Infusion) made by C. R. Bard, Inc., cleared under FDA 510(k) K883577 on 18 Nov 1988, active as of 22 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK883577 | Date18 Nov 1988 | StatusActive |
Official records
- FDA 510(k) database K883577
From the FDA record
- Received by the FDA
- 22 Aug 1988
- FDA decision
- 18 Nov 1988 (Substantially Equivalent)
- FDA review time
- 88 days from receipt to decision
Recalls
-
- Terminated Z-1378-2011
Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the InfusOR will not operate. The reason being that the InfusOR (2L3100) Battery Spring has a flat contact that measures .53 inches .01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum
Timeline
- Cleared FDA 510(k) K883577
- Recall initiated FDA recalls Z-1378-2011
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- FRN
- FDA applicant
- C.R. Bard, Inc.
- Device class
- Class II
- Regulation
- 21 CFR Part 880.5725
- Advisory committee
- General Hospital
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code FRN (Pump, Infusion).
- MODEL 210 INFUSION PUMP by Avi, Inc.
- ABBOTT XL3M INFUSION PUMP by Abbott Laboratories
- MODEL 285 INFUSION PUMP, MODIFICATION by 3M Health Care, Ltd.
- INTERMEDICS SYRINGE PUMP 710-02 by Intermedics, Inc.
- BURRON AMBULATORY DRUG DELIVERY SYSTEM by Burron Medical, Inc.
- SIGNATURE EDITION INFUSION PUMP, SIGNATURE EDITION ADMINISTRATION SETS, MODELS 70XX, 71XX, 72XX, 7XXXX SERIES by Alaris Medical Systems, Inc.
- SYRINGE DRIVER 209 by Medix Medical Electronics (U.S.A.), Inc.
- ANSPACH IRRIGATION PUMP AND IRRIGATION PUMP II by The Anspach Effort, Inc.
- PUMPMAN(TM) AMBULATORY DRUG DELIVERY SYSTEM by Strato Medical Corp.
- PROVIDER U-100 INSULIN INFUSION SYS by Pancretec, Inc.