Bard Mini-Infuser 1500 & 3000
510(k) K930459
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK930459 | Date10 Aug 1993 | StatusActive |
Recalls
-
- Terminated Z-2269-2012
- Terminated Z-2268-2012
Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches .01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FRN
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5725
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 26 Sep 2026
The Bard Mini-Infuser 1500 & 3000 is an infusion pump designed for the controlled administration of intravenous medications. It functions as a syringe pump to deliver fluids at regulated rates.
This device is categorized under the General Hospital medical specialty and has received FDA 510(k) clearance. It is identified by the regulation number 880.5725 and product code FRN.