Bayer Advia 1650 Chemistry System

by Bayer Corp.

510(k) K990346

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

14 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK990346 Date21 May 1999ClearedStatusActive

Recalls

14 recalls · FDA

    • Terminated Z-2014-2021 FDA Class II initiated 18 May 2021

    ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously elevated or depressed patient result to be reported without alert on the test result

    • Terminated Z-0755-2016 FDA Class II initiated 7 Oct 2015 on shared 510(k) K990346

    Multiple Software issues. Auto Start-UP Fail, Calibration Interval Resets when a Reagent Blank is run, Control Definition screen assumes range defined is 2 SD, Laboratory Information System (LIS) communication/Laboratory Automation (LAS) issue, Printer Driver Resets, ISE Calibration Ranges are too conservative for Urine Sodium, Archiving and deletion may fail and Workstation services may restart.

    • Terminated Z-0183-2016 FDA Class II initiated 28 Aug 2015

    ISE module may produce discrepant, unflagged results for Sodium, Potassium or Chloride when the respective electrode is left unplugged after replacement, washing or maintenance activities. The calibration and 2-level QC will fail if the electrode is left unplugged. The IFU states to calibrate the ISEs after replacement and to run two levels of controls before running patient samples.

    • Terminated Z-1399-2015 FDA Class II initiated 11 Feb 2015

    Siemens Healthcare Diagnostics has confirmed an issue with ADVIA Chemistry XPT software version 1.0.2. The system may cause samples to remain in an Inprocess state. Test results on a sample that is held Inprocess will not transmit to the LIS. Manual intervention is necessary to complete the processing of the samples that are held Inprocess.

    • Terminated Z-0800-2014 FDA Class II initiated 17 Dec 2013 on shared 510(k) K990346

    An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.

    • Terminated Z-0940-2014 FDA Class II initiated 6 Nov 2013 on shared 510(k) K990346

    The ADVIA Chemistry software systems using certain software versions allows the system to automatically run assay calibrations and quality control (QC).

  • 2 recalls, same reason

    • Terminated Z-0343-2014 FDA Class II initiated 3 Sep 2013
    • Terminated Z-0344-2014 FDA Class II initiated 3 Sep 2013

    Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or 2400 Chemistry Systems.

  • 2 recalls, same reason

    • Terminated Z-0353-2007 FDA Unclassified (pre-2012) initiated 8 Sep 2006 on shared 510(k) K990346
    • Terminated Z-0352-2007 FDA Unclassified (pre-2012) initiated 8 Sep 2006 on shared 510(k) K990346

    Bayer received a customer complaint where the Valproic Acid Quality control material recovered lower than target when it was run in a panel of other Therapeutic Drug (TDM) assays. The under recovery observed by the customer was approximately -25%. Investigation into this issue was initiated promptly following receipt of this complaint.

    • Terminated Z-0351-2007 FDA Unclassified (pre-2012) initiated 8 Sep 2006 on shared 510(k) K990346

    Bayer received a customer complaint where the Valproic Acid Quality control material recovered lower than target when it was run in a panel of other Therapeutic Drug (TDM) assays. The under recovery observed by the customer was approximately -25%.

  • 3 recalls, same reason

    • Terminated Z-1448-06 FDA Unclassified (pre-2012) initiated 17 Feb 2006
    • Terminated Z-1449-06 FDA Unclassified (pre-2012) initiated 17 Feb 2006
    • Terminated Z-1450-06 FDA Unclassified (pre-2012) initiated 17 Feb 2006

    Low QC recovery observed on carbamazepine (CARB) results immediately following a Gentamicin (GENT) assay or a Digoxin (DIG) assay. When GENT or DIG precedes CARB, the CARB test result is artificially low (as much as -25%). PHNY was determined to also exhibit a similar carryover effect on CARB. All others were found not to produce the same reagent probe-based carryover effect.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JGS
Device Class
Class II
Regulation
21 CFR Part 862.1665
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent