Bayer Rapidpoint 400 System
by Bayer Corp.
510(k) K002738
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK002738 | Date19 Dec 2000 | StatusActive |
Recalls
-
- Ongoing Z-2230-2025
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
-
- Terminated Z-2912-2016
Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge
-
- Terminated Z-2797-2016
There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system
-
- Terminated Z-2479-2015
When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to include either: duplicate data, missing data, or data from a different patient.
-
- Terminated Z-0102-2014
RAPIDPoint 400 Series, RAPIDPoint 500, or RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
-
- Terminated Z-1019-2010
- Terminated Z-1020-2010
Software: Reference sensor failure may result in inaccurate pH reuslts
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CHL
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1120
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent