Bci 3180 Pulse Oximeter
by Bci, Inc.
510(k) K020350
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK020350 | Date30 Apr 2002 | StatusActive |
Recalls
-
- Terminated Z-0919-2010
Smiths Medical is conducting a voluntary recall of a limited number of medical device AC power cords, manufactured by Electri-Cord Manufacturing Co., for use with BCI 3180 Pulse Oximeter(Human), 9200 Advisor Vital Signs Monitor(Human or Veterinary), 8400 Capnocheck II Capnometer(Human or Veterinary) and V6400 Invasive Pressure Monitor (Veterinary). The affected Electri-Cord AC power cords are eq
-
- Terminated Z-1619-2008
The Smith Medical PM, Inc. BCI 3180 Pulse Oximeter tantalum capacitors C38, C72, C74, and C98 were installed backwards during assembly at the board manufacturer. This exposes the components to reverse polarity voltages that may lead to their pre-mature failure. As a result, the identified monitors may exhibit failure modes such as the monitor may shut down and not produce alarms or signals to
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DQA
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.2700
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent