Skip to content

Bci 3180 Pulse Oximeter

by Bci, Inc.

510(k) K020350

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK020350 Date30 Apr 2002ClearedStatusActive

Recalls

2 recalls · FDA

    • Terminated Z-0919-2010 FDA Unclassified (pre-2012) initiated 15 Jan 2010

    Smiths Medical is conducting a voluntary recall of a limited number of medical device AC power cords, manufactured by Electri-Cord Manufacturing Co., for use with BCI 3180 Pulse Oximeter(Human), 9200 Advisor Vital Signs Monitor(Human or Veterinary), 8400 Capnocheck II Capnometer(Human or Veterinary) and V6400 Invasive Pressure Monitor (Veterinary). The affected Electri-Cord AC power cords are eq

    • Terminated Z-1619-2008 FDA Unclassified (pre-2012) initiated 27 Feb 2008

    The Smith Medical PM, Inc. BCI 3180 Pulse Oximeter tantalum capacitors C38, C72, C74, and C98 were installed backwards during assembly at the board manufacturer. This exposes the components to reverse polarity voltages that may lead to their pre-mature failure. As a result, the identified monitors may exhibit failure modes such as the monitor may shut down and not produce alarms or signals to

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DQA
Device Class
Class II
Regulation
21 CFR Part 870.2700
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Bci, Inc.