Bcs System
by Dade Behring, Inc. 510(k)
Active
510(k) K013114
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK013114 | Date9 Nov 2001 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0824-2016
False short clotting times for PT on BCS and BCS XP with Dade Innovin can occur. This has also an effect for false low PT INR values. These false short clotting times were observed only with turbid, hemolytic and/or icteric samples from intensive care patients. The corresponding reaction curves of these samples showed a pre-peak, which can lead to falsely short clotting time. The result was not flagged by the existing check algorithms.
-
- Terminated Z-1867-2012
Firm has confirmed that if the Cleaner SCS vial becomes empty after processing a test for Antithrombin assay using the INNOVANCE(R) Antithrombin kit, and the test immediately following is an APTT based clotting assay several results may be shortened.
Specification
- 510(k) Number
- FDA Product Code
- JPA
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.5425
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent